动态电流聚焦与单极刺激在人工耳蜗使用者居家试验中的比较
本研究在13例人工耳蜗使用者中比较动态电流聚焦(DCF)与单极刺激(MP)的言语识别、频谱时间与时间处理表现,9例完成单盲自身交叉试验。DCF在65 dB SPL下使SMRT阈值改善0.40 ripples/octave(p=0.02),但Bonferroni校正后p=0.1;DCF平均电池寿命缩短48%(5.1小时)(p<0.001)。作者认为DCF证据薄弱,未一致改善噪声下言语识别。
要点
- DCF在65 dB SPL下SMRT阈值改善0.40 ripples/octave,但Bonferroni校正后未达显著。
- DCF与MP在噪声下言语识别、时间测试、STRIPES及45 dB SPL SMRT上无显著差异。
- DCF平均电池寿命较MP缩短48%(5.1小时)(p<0.001)。
核心结论
- DCF可能在舒适响度(65 dB SPL)下改善人工耳蜗使用者的频谱分辨率,但本研究证据薄弱且Bonferroni校正后显著结果消失。
- 并非所有频谱测试均显示该改善,DCF电池寿命如预期缩短。
- 作者指出,尽管本研究受小样本量限制,结合既往研究,DCF并未较MP策略一致改善噪声下言语识别。
研究设计与样本
- 单盲自身对照交叉试验,13例人工耳蜗使用者验配DCF与非转向MP(脉冲速率、脉宽和活动电极匹配),9例完成试验。
- 重复测试共四次,策略分配为DCF-MP-DCF-MP或MP-DCF-MP-DCF,每段适应期3周,结束时进行测试。
- 测试包括矩阵噪声下言语识别、频谱时间纹波测试(SMRT和STRIPES)及时间幅度调制检测,所有参与者用言语、空间与听觉质量量表记录主观体验。
主要发现
- DCF较MP在65 dB SPL下SMRT阈值改善0.40 ripples/octave(p=0.02,Bonferroni校正后p=0.1)。
- 两种策略在对话响度(65 dB SPL)与柔和响度(45 dB SPL)噪声下言语识别、时间测试、STRIPES及45 dB SPL SMRT上均无显著差异。
- Bonferroni校正后,矩阵噪声下言语识别测试在两种响度下均存在学习效应(65 dB SPL:p=0.01;45 dB SPL:p=0.02),DCF与MP随时间的学习效应无差异。
- 言语、空间与听觉质量量表主观体验无显著差异。
- DCF较MP平均电池寿命缩短48%(5.1小时)(p<0.001)。
局限性
- 本研究受小样本量限制。
- 本研究证据薄弱,Bonferroni校正后显著结果消失。
- 并非所有频谱测试均显示改善。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: This study compared the performance of a dynamic partial tripolar cochlear implant speech encoding strategy termed dynamic current focusing (DCF) to monopolar stimulation (MP) using spectro-temporal, temporal, and speech-in-noise recognition testing. DESIGN: DCF is a strategy that utilizes tripolar or high partial tripolar stimulation at threshold level and increases loudness by slowly widening current spread towards most comfortable level. Thirteen cochlear implant users were fitted with DCF and a non-steered MP matched on pulse rate, pulse width, and active electrodes. Nine participants completed the single-blinded within-subject crossover trial. Repeated testing consisted of four sessions. Strategies were allocated in a DCF-MP-DCF-MP or MP-DCF-MP-DCF design. Three-week adaptation periods ended with a test session in which speech-in-noise recognition (matrix speech-in-noise sentence test), spectro-temporal ripple tests (SMRT and STRIPES) and a temporal amplitude modulation detection test were conducted. All participants recorded their subjective experiences with both strategies using the Speech, Spatial and Qualities of Hearing Scale questionnaire. RESULTS: Participants' SMRT thresholds improved 0.40 ripples per octave ( p = 0.02, Bonferroni-corrected: p = 0.1) with DCF over MP at 65 dB SPL. No significant differences between the strategies were found on speech-in-noise recognition at conversational (65 dB SPL) and soft (45 dB SPL) loudness levels, temporal testing, STRIPES, or the SMRT at 45 dB SPL. After Bonferroni correction, a learning effect remained on the matrix speech-in-noise sentence test at both loudness levels (65 dB SPL: p = 0.01; 45 dB SPL: p = 0.02). There was no difference in learning effects over time between DCF and MP. Similarly, no significant differences were found in subjective experience on the Speech, Spatial and Qualities of Hearing Scale questionnaire. DCF reduced average battery life by 48% (5.1 hours) ( p < 0.001) compared to MP. CONCLUSIONS: DCF may improve spectral resolution over MP at comfortable loudness (65 dB SPL) in cochlear implant users. However, the evidence collected in this study was weak and the significant result disappeared after Bonferroni correction. Also, not all spectral tests revealed this improvement. As expected, battery life was reduced for DCF. Although the current study is limited by its small sample size, considering previous studies, DCF does not consistently improve speech recognition in noise over MP strategies.
引用
- van Groesen NRA, Briaire JJ, de Jong MAM, et al. Dynamic Current Focusing Compared to Monopolar Stimulation in a Take-Home Trial of Cochlear Implant Users. Ear Hear. 2022. PMID: 36279119. doi:10.1097/AUD.0000000000001282
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。