脉冲波形与相间间隙对电听觉言语感知和主观音质的影响
一项前瞻性单次访视研究纳入20例无面神经刺激的成年人工耳蜗使用者,比较双相与三相脉冲及2.1、10、20微秒相间间隙的 fitting 图。结果显示三相刺激和更长相间间隙的言语感知与临床图相似,但三相脉冲在急性测试中多项主观音质评分较低;作者认为三相脉冲不影响噪声下言语感知,可作为人工耳蜗调机的可选方案。
要点
- 三相刺激与临床图相比噪声下言语感知相似(P=0.891)。
- 更长相间间隙与临床图相比言语感知相似(P=0.361)。
- 三相脉冲在急性测试中多项主观音质评分低于双相脉冲。
核心结论
- 三相脉冲刺激不影响噪声下言语感知,尽管急性测试中部分音质属性评分低于临床图,仍可作为人工耳蜗调机的有价值选项。
研究设计与样本
- 前瞻性单次访视研究,纳入20例无面神经刺激的成年人工耳蜗使用者。
- 基于受试者当前临床调机图,创建6种不同脉冲波形(双相和三相)及相间间隙(2.1微秒、10微秒、20微秒)的调机图。
- 先通过调整刺激电荷量使各配置响度平衡,再用德语矩阵句子测试(奥尔登堡句子测试)测量噪声下言语感知,并用视觉模拟量表评估语音和音乐特定音质属性及总体偏好。
主要发现
- 三相刺激与受试者临床调机图设置相比,噪声下言语感知相似(P=0.891)。
- 更长相间间隙与临床调机图设置相比,噪声下言语感知相似(P=0.361)。
- 相间间隙恒定时,三相刺激达到等响所需的刺激幅度显著高于双相刺激。
- 增加相间间隙对感知响度的影响(P<0.0001)以及三相脉冲相比双相脉冲在等响时电荷减少的影响显著更大。
- 在无适应时间的急性设置中,三相配置在语音可懂度、清晰度、自然度和总体偏好以及音乐自然度和总体偏好方面的主观音质评分低于双相。
临床怎么用
- 作者认为三相脉冲刺激不影响噪声下言语感知,可作为人工耳蜗调机的有价值选项。
局限性
- 急性测试设置中部分音质属性评分低于临床调机图。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Stimulation with triphasic pulses has been shown to reduce the occurrence of unwanted facial nerve stimulation (FNS) with cochlear implants (CIs). However, there is little data available on how different pulse shapes affect the hearing outcome with electrical hearing in general. The aim of the study was to evaluate the effects of different stimulation pulse shapes on speech perception in noise, as well as loudness perception and subjective sound quality. METHODS: Twenty experienced cochlear-implant users not suffering from FNS participated in a prospective single-visit study. Based on the subjects' current clinical fitting, six fitting maps with different pulse shapes (biphasic and triphasic) and different interphase gap (IPG) durations (2.1 µs, 10 µs, and 20 µs) were created. First, the loudness was balanced for each configuration by adjusting the stimulation charge amount. Then, speech perception in noise was measured with a German matrix sentence test (Oldenburg Sentence test). The perception of particular sound attributes of speech and music, as well as overall preference, was evaluated with visual analog scales. RESULTS: Similar levels of speech perception were obtained with triphasic stimulation ( P = 0.891) and longer IPGs ( P = 0.361) compared to the subjects' clinical map settings. The stimulation amplitudes for equal loudness were significantly higher with triphasic stimulation compared to biphasic stimulation when keeping the IPG constant. Increasing the IPG had a significantly larger effect on perceived loudness ( P < 0.0001) and charge reduction for equal loudness with triphasic pulses compared to biphasic pulses. Triphasic configuration showed lower overall subjective sound quality ratings than biphasic for speech intelligibility, clarity, naturalness, and overall preference, as well as for music naturalness, and overall preference in the acute setting without adaptation time. Post-hoc pairwise comparisons against the clinical map revealed significantly lower speech naturalness ratings for triphasic with 2.1 µs IPG and for triphasic with 20 µs IPG only. CONCLUSION: Although some sound quality attributes were rated lower compared to the clinical map in the acute test setting, stimulation with triphasic pulses does not affect speech perception in noise and can be considered as a valuable option in CI fitting.