文献速递耳科与耳神经小儿耳鼻咽喉

单侧与双侧耳聋人工耳蜗使用者中声学变化复合波与听力学表现的比较

The Acoustic Change Complex Compared to Hearing Performance in Unilaterally and Bilaterally Deaf Cochlear Implant Users.
PubMed 文献 · Ear Hear · 2022-10-18

研究纳入13例双侧耳聋和6例单侧耳聋、均单侧使用人工耳蜗至少1年的受试者,比较由频率变化诱发的声学变化复合波(ACC)与言语感知及频率辨别阈的关系。结果显示17/19例可记录到可重复ACC,最大频率变化(+3电极步)记录成功率最高;人工耳蜗耳的N1潜伏期和N1-P2波幅与言语感知及频率辨别相关,单侧耳聋者人工耳蜗耳较对侧正常听力耳潜伏期更长、波幅更小。作者认为ACC可用于无法可靠完成行为测试者的听力评估。

要点

  • 17/19例受试者可记录到可重复的ACC,最大频率变化(+3电极步)记录成功率最高。
  • 人工耳蜗耳的N1潜伏期和N1-P2波幅与言语感知及频率辨别相关。
  • 单侧耳聋者人工耳蜗耳较对侧正常听力耳潜伏期更长、波幅更小。

核心结论

  • 由纯音频率变化诱发的ACC潜伏期和波幅与人工耳蜗使用者的频率辨别及言语感知能力相关性良好。
  • 对于无法可靠完成行为测试的患者,ACC可能在评估听力表现方面具有附加价值。

研究设计与样本

  • 纳入13例双侧耳聋和6例单侧耳聋受试者,所有受试者均单侧使用人工耳蜗至少1年。
  • 言语感知采用辅音-元音-辅音测试(+10 dB信噪比)和数字噪声测试评估。
  • 频率辨别阈在两种参考频率下测量,采用3间隔、2备选强制选择的自适应阶梯程序。
  • 两种参考频率依据每名受试者的频率分配表选取,分别位于包含500 Hz或2000 Hz的电极频带中心,对应蜗顶电极或中间电极。
  • ACC由与上述两种参考频率相同的纯音诱发,频率增加幅度包括:在中间或蜗顶电极频带内(+0.25电极步),以及至第1(+1)、第2(+2)和第3(+3)相邻电极中心频率的步进。

主要发现

  • 19例受试者中17例记录到可重复的ACC,最大频率变化(+3电极步)时记录成功率最高。
  • 频率变化越大,N1潜伏期越短、N1-P2波幅越大。
  • 在单侧和双侧耳聋受试者中,人工耳蜗耳的N1潜伏期和N1-P2波幅均与言语感知及频率辨别相关,即短潜伏期和大波幅提示更好的言语感知和频率辨别。
  • 单侧耳聋者与双侧耳聋者的人工耳蜗耳之间,ACC潜伏期和波幅无显著差异。
  • 单侧耳聋者的人工耳蜗耳较其对侧正常听力耳潜伏期明显更长、波幅更小。

临床怎么用

  • 作者认为,对于无法可靠完成行为任务的患者,ACC可在评估听力表现方面具有附加价值。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Clinical measures evaluating hearing performance in cochlear implant (CI) users depend on attention and linguistic skills, which limits the evaluation of auditory perception in some patients. The acoustic change complex (ACC), a cortical auditory evoked potential to a sound change, might yield useful objective measures to assess hearing performance and could provide insight in cortical auditory processing. The aim of this study is to examine the ACC in response to frequency changes as an objective measure for hearing performance in CI users. DESIGN: Thirteen bilaterally deaf and six single-sided deaf subjects were included, all having used a unilateral CI for at least 1 year. Speech perception was tested with a consonant-vowel-consonant test (+10 dB signal-to-noise ratio) and a digits-in-noise test. Frequency discrimination thresholds were measured at two reference frequencies, using a 3-interval, 2-alternative forced-choice, adaptive staircase procedure. The two reference frequencies were selected using each participant's frequency allocation table and were centered in the frequency band of an electrode that included 500 or 2000 Hz, corresponding to the apical electrode or the middle electrode, respectively. The ACC was evoked with pure tones of the same two reference frequencies with varying frequency increases: within the frequency band of the middle or the apical electrode (+0.25 electrode step), and steps to the center frequency of the first (+1), second (+2), and third (+3) adjacent electrodes. RESULTS: Reproducible ACCs were recorded in 17 out of 19 subjects. Most successful recordings were obtained with the largest frequency change (+3 electrode step). Larger frequency changes resulted in shorter N1 latencies and larger N1-P2 amplitudes. In both unilaterally and bilaterally deaf subjects, the N1 latency and N1-P2 amplitude of the CI ears correlated to speech perception as well as frequency discrimination, that is, short latencies and large amplitudes were indicative of better speech perception and better frequency discrimination. No significant differences in ACC latencies or amplitudes were found between the CI ears of the unilaterally and bilaterally deaf subjects, but the CI ears of the unilaterally deaf subjects showed substantially longer latencies and smaller amplitudes than their contralateral normal-hearing ears. CONCLUSIONS: The ACC latency and amplitude evoked by tone frequency changes correlate well to frequency discrimination and speech perception capabilities of CI users. For patients unable to reliably perform behavioral tasks, the ACC could be of added value in assessing hearing performance.