文献速递耳科与耳神经

人工耳蜗植入后适应恢复时间延长的语后聋成人听神经水平倾向于较差言语感知表现

Postlingually Deafened Adult Cochlear Implant Users With Prolonged Recovery From Neural Adaptation at the Level of the Auditory Nerve Tend to Have Poorer Speech Perception Performance.
PubMed 文献 · Ear Hear · 2022-10-18

本研究纳入18名语后聋成人人工耳蜗使用者,用电诱发复合动作电位评估听神经适应与适应恢复,发现适应恢复速度与安静及噪声下CNC词得分呈中等负相关,分别解释14.1%和16.7%的变异,适应恢复延长者言语感知表现倾向更差。

要点

  • 适应恢复速度与安静下CNC词得分呈中等负相关,解释14.1%变异。
  • 适应恢复速度与噪声下CNC词得分呈中等负相关,解释16.7%变异。
  • 适应指数、适应恢复比与适应速度同词得分的相关性为可忽略至弱。

核心结论

  • 听神经适应恢复速度较其他适应与恢复特征对语后聋成人人工耳蜗使用者言语感知更重要,适应恢复延长者言语感知表现倾向更差。

研究设计与样本

  • 研究纳入18名语后聋成人,均为Cochlear Nucleus装置使用者且测试耳为全电极阵列植入。
  • 用电诱发复合动作电位(eCAP)评估听神经适应与适应恢复。
  • 适应量以三个时间窗的适应指数量化:0至8.89 ms(窗1)、44.44至50.00 ms(窗2)、94.44至100.00 ms(窗3)。
  • 适应恢复以掩蔽-探测间隔1.054至256 ms对数步进记录100 ms脉冲串最后脉冲的eCAP,用适应恢复比量化,时间常数用最多三成分指数函数估计。
  • 言语感知用安静及信噪比+10 dB言语形噪声下的辅音-核-辅音(CNC)词得分评估,采用单尾Pearson积矩相关与多变量线性回归分析。

主要发现

  • 适应恢复速度与安静下CNC词得分呈中等负相关,解释14.1%的变异。
  • 适应恢复速度与噪声下CNC词得分呈中等负相关,解释16.7%的变异。
  • CNC词得分与适应指数、适应恢复比及适应速度的相关强度为可忽略至弱。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: This study investigated the effects of two temporal response properties of the auditory nerve (i.e., neural adaptation and recovery from neural adaptation) on speech perception performance in postlingually deafened adult cochlear implant (CI) users. DESIGN: Study participants included 18 postlingually deafened adults who were Cochlear Nucleus device users with a full electrode array insertion in the test ear(s). Neural adaptation and adaptation recovery of the auditory nerve (AN) were evaluated using electrophysiological measures of the electrically evoked compound action potential (eCAP). The amount of neural adaptation was quantified by the adaptation index within three time windows: 0 to 8.89 (window 1), 44.44 to 50.00 (window 2), and 94.44 to 100.00 ms (window 3). The speed of neural adaptation was estimated using a two-parameter power law function. To evaluate adaptation recovery of the AN, eCAPs to the last pulse of the 100-ms pulse train were recorded at masker-probe-intervals ranging from 1.054 to 256 ms in logarithmic steps. The amount of adaptation recovery was quantified by the adaptation recovery ratio. The time-constant of adaptation recovery was estimated using an exponential function with up to three components. Speech perception performance was evaluated by measuring consonant-nucleus-consonant (CNC) word scores presented in quiet and in speech-shaped noise at a signal-to-noise ratio (SNR) of +10 dB. One-tailed Pearson Product Moment correlation tests were used (1) to assess the associations among parameters of neural adaptation and adaptation recovery and (2) to evaluate the strength of association between these parameters and CNC word scores measured in quiet and in noise. The contributions of different parameters quantifying neural adaptation and adaptation recovery on speech perception scores were evaluated using multivariable linear regression analyses. RESULTS: The Pearson Product Moment correlation coefficient demonstrated a moderate, negative correlation between the speed of adaptation recovery and CNC word scores measured in quiet and in noise. The speed of adaptation recovery accounted for 14.1% of variability in CNC word scores measured in quiet and 16.7% of variability in CNC word scores measured in noise. The correlation strengths between CNC word scores and the adaptation index, the adaptation recovery ratio and the speed of neural adaptation ranged from negligible to weak. CONCLUSIONS: The speed of adaptation recovery plays a more important role than other features of neural adaptation and adaptation recovery of the AN in speech perception in postlingually deafened adult CI users. Patients with prolonged adaptation recovery tend to show poorer speech perception performance.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. He S, Skidmore J, Carter BL, et al. Postlingually Deafened Adult Cochlear Implant Users With Prolonged Recovery From Neural Adaptation at the Level of the Auditory Nerve Tend to Have Poorer Speech Perception Performance. Ear Hear. 2022. PMID: 35652833. doi:10.1097/AUD.0000000000001244

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。