文献速递耳科与耳神经小儿耳鼻咽喉

儿童人工耳蜗使用者中单极电凝的使用

Monopolar Cautery Use in Pediatric Cochlear Implant Users.
PubMed 文献 · Otolaryngol Head Neck Surg · 2023-01-27

该研究通过回顾性病例系列和系统综述,评估儿童人工耳蜗使用者后续手术中使用单极电凝的发生率及对植入装置功能的影响。单中心2012至2021年190例患儿中,15例(7.9%)在17次手术中使用单极电凝,术后无装置故障或性能下降;系统综述另发现4例类似患者,亦无功能改变。

要点

  • 儿童人工耳蜗使用者后续手术中单极电凝使用率为7.9%(15/190)。
  • 使用单极电凝的17次手术中,7次涉及头颈部,10次位于锁骨以下。
  • 术后无装置故障或人工耳蜗性能下降,系统综述补充4例亦无功能改变。

核心结论

  • 单极电凝不一定损伤人工耳蜗功能,但最规避风险策略仍是避免使用单极电凝。

研究设计与样本

  • 回顾性病例系列,来自单中心三级学术转诊中心2012至2021年儿童人工耳蜗患者病历,并辅以系统文献综述。

主要发现

  • 190例患儿后续手术中,15例(7.9%)在17次手术中使用单极电凝,其中7次(41.2%)涉及头颈部,10次位于锁骨以下。
  • 使用单极电凝后无患者出现装置故障或人工耳蜗性能下降。
  • 系统综述额外识别4例人工耳蜗植入后使用单极电凝手术的患者,未发现人工耳蜗功能改变。

临床怎么用

  • 作者认为单极电凝不应立即与植入装置功能障碍关联,但规划手术时最规避风险的策略是避免使用单极电凝。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To determine the incidence and impact of monopolar cautery use in a cohort of pediatric cochlear implant (CI) users. STUDY DESIGN: Case series from a retrospective chart review and a systematic review of the literature. SETTING: Tertiary academic referral center. METHODS: CI patient charts from 2012 to 2021 were reviewed from a single pediatric hospital system to determine if monopolar cautery was used during a subsequent surgical procedure. In addition, a systematic review of the literature was performed to identify additional, relevant patients. Postoperative CI function was the primary outcome measure. RESULTS: In total, 190 patients underwent a surgical procedure following cochlear implantation in a single pediatric hospital system. Fifteen patients (7.9%) and 17 distinct surgical procedures were identified in which monopolar cautery was used. Seven of these 17 cases (41.2%) involved the head and neck, and 10 were performed below the clavicles. No patients experienced a device failure or a decline in CI performance following surgery. A systematic review identified an additional 4 patients who underwent a surgery that used monopolar cautery following cochlear implantation, and no change in CI function was identified. CONCLUSIONS: The present study adds additional support to the notion that monopolar cautery does not necessarily injure CI functionality. While the most risk adverse strategy when planning a surgical procedure for a CI patient is to avoid monopolar cautery use altogether, the use of cautery should not immediately be associated with implant dysfunction.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。

引用

  1. Page JC, Chapel AC, Silva RC, et al. Monopolar Cautery Use in Pediatric Cochlear Implant Users. Otolaryngol Head Neck Surg. 2023. PMID: 35763369. doi:10.1177/01945998221108051

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。