文献速递耳科与耳神经

通过人工耳蜗直接记录人工耳蜗植入者的皮层听觉诱发电位:一项可行性研究

Cortical Auditory Evoked Potentials Recorded Directly Through the Cochlear Implant in Cochlear Implant Recipients: a Feasibility Study.
PubMed 文献 · Ear Hear · 2022-03-03

本研究评估通过人工耳蜗直接、单独记录皮层听觉诱发电位(CAEP)的可行性。9例人工耳蜗植入者(7例语后聋成人,2例语前聋儿童)对侧耳有残余听力,经非植入耳给声,在植入耳记录CAEP。结果显示经人工耳蜗记录的CAEP与头皮电极同步记录的标准CAEP等效,以盒式电极作参考时形态、潜伏期和波幅高度相关。

要点

  • 经人工耳蜗直接记录的CAEP与头皮电极同步记录的标准CAEP等效。
  • 以盒式电极作参考时,CAEP形态、潜伏期和波幅高度相关。
  • 2例儿童CAEP潜伏期与正常发育年龄组一致,且较成人延迟,符合预期。

核心结论

  • 本研究证明在具有残余听力的成人中,通过人工耳蜗直接、单独记录长潜伏期CAEP是可行的。
  • 经人工耳蜗记录的CAEP与外部诱发电位系统经头皮电极同步记录的CAEP高度相似。
  • 无需外部记录系统而直接经植入体记录的能力,是一种具有许多临床和研究意义的新方法。

研究设计与样本

  • 可行性研究,共纳入9例人工耳蜗植入者,包括4例女性和5例男性。
  • 其中7例为语后聋成人(23至72岁),2例为语前聋儿童(1例青少年15岁,1例儿童8岁)。
  • 所有参与者人工耳蜗对侧耳均有残余听力。
  • 经非植入耳给予舒适聆听水平的短纯音刺激,在植入耳记录CAEP。
  • 记录采用颅内蒙太奇,以耳蜗内尖部电极作为有源电极,以两个耳蜗外植入电极(盒式电极和环式电极)之一作为参考电极。

主要发现

  • 经人工耳蜗直接、单独记录的CAEP与标准头皮记录的CAEP等效。
  • 以盒式电极作参考时,CAEP成分的形态、潜伏期和波幅高度相关。
  • 2例儿童患者的CAEP潜伏期与其正常发育年龄组一致,且较成人CAEP潜伏期延迟,符合预期。

关键数值与推荐

  • 以盒式电极作参考时,CAEP成分的形态、潜伏期和波幅高度相关。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The use of objective measures in cochlear implant (CI) mapping, has greatly contributed to the refinement of the setting of audible and comfortable stimulation levels, which serve as the basis of the mapping process, especially in cases of infants and young children. In addition, objective measures can also confirm the integrity of the CI system. Current CI objective measures mainly reflect neural activity from the auditory nerve and brainstem site. An objective cortical CI measure that reflects directly central auditory activity is greatly needed, especially since it is closely related to CI outcomes in both children and adults. Recording the brain activity currently requires an external evoked potential (EP) system including scalp electrodes, rendering it impractical for widespread clinical use. This study aimed to assess the feasibility of recording cortical auditory evoked potentials (CAEPs) directly and solely through the cochlear implant in response to external acoustic stimulation in the non-implanted ear. DESIGN: A total of nine CI users (four females and five males) participated, including seven post-lingual adults (23 to 72 years), and two pediatric cases, one teenager (15 years), and one child (8 years)-both pre-lingual. All participants had a residual hearing in the ear contralateral to the ear with CI. CAEPs were recorded in the implanted ear in response to acoustic stimulation of the non-implanted ear, consisting of a brief tonal stimulus at comfortable listening levels. Recordings used an intracranial montage consisting of an intracochlear apical electrode (active) and one of the two ( case and ring ) extra-cochlear implanted electrodes serving as reference electrodes. The CI CAEPs were compared with a single-channel conventional CAEP recording obtained simultaneously via scalp electrodes (Fz-mastoid) using a standard EP system and an external trigger from the CI system. Statistical comparisons were made between the CI and the scalp recorded CAEPs and for differences between the CI CAEP measures acquired using the ring and the case as the reference electrode. RESULTS: CAEPs recorded directly and solely through the CI were equivalent to the standard scalp recorded CAEP responses. CAEP responses acquired using the case electrode as the reference were highly correlated in terms of morphology, latencies, and amplitudes of the CAEP components. The CI CAEP latencies of the two pediatric cases were consistent with their normal developed age group and delayed relative to adult CAEP latencies, as expected. CONCLUSIONS: This study demonstrated the feasibility of recording long latency CAEPs directly and solely through CI in adults with residual hearing, in response to acoustic stimulation of the non-implanted ear. The CI CAEPs closely resembled the CAEPs recorded simultaneously by an external EP system and via scalp electrodes. The ability to record directly from the implant, without the need of an external recording system, presents an innovative method with many clinical and research implications.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Attias J, HabibAllah S, Aditya Tarigoppula VS, et al. Cortical Auditory Evoked Potentials Recorded Directly Through the Cochlear Implant in Cochlear Implant Recipients: a Feasibility Study. Ear Hear. 2022. PMID: 35245922. doi:10.1097/AUD.0000000000001212

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。