Omnipod胰岛素泵中二丙二醇二丙烯酸酯致变应性接触性皮炎
本文报告3例Omnipod胰岛素泵所致变应性接触性皮炎(ACD)患者,并通过气相色谱-质谱联用分析不同批次泵的化学成分。
要点
- Omnipod泵的成分随时间变化,斑贴试验需使用更新后的系列。
- 长期使用医疗器械的儿童可能发生多种接触过敏及交叉过敏。
主要发现
- 患者所带Omnipod泵提取物中检出DPGDA。
- 早期批次Omnipod泵含三丙二醇二丙烯酸酯、IBOA、N,N-二甲基丙烯酰胺、二乙二醇乙醚丙烯酸酯(DEGEA),但不含DPGDA。
- 1例患者除未检测的DEGEA外,对上述其他过敏原均呈阳性反应。
局限性
- 仅报告3例病例,样本量小;未系统评估儿童人群及长期使用器械的过敏风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Cases of allergic contact dermatitis (ACD) caused by isobornyl acrylate (IBOA) in the Omnipod® insulin pump have previously been reported. OBJECTIVES: To present three cases of patients with ACD caused by a new allergen in the pump, and results from chemical analyses. METHODS: Omnipod pumps from different batches were analysed by gas chromatography-mass spectrometry. Aimed testing, with the department's medical device (MD) series and substances identified in the pump including dipropylene glycol diacrylate (DPGDA) at 0·01% and 0·1% in petrolatum (pet.), was performed. Patch testing also included extracts from the device, the adhesive patch as is, and allergens from baseline series. RESULTS: All patients tested positive to 0·1% DPGDA in pet., and two patients additionally to a 0·01% concentration. DPGDA was found in extracts of the Omnipod pumps brought by the patients. An Omnipod pump from an earlier batch contained tripropylene glycol diacrylate, IBOA, N,N-dimethylacrylamide, di(ethylene glycol)ethyl ether acrylate (DEGEA) but no DPGDA. One of the patients reacted positively to all of these allergens except DEGEA, which was not tested. CONCLUSIONS: When suspecting ACD to MDs, DPGDA at 0·1% in pet. should be tested. The contents of Omnipod have changed over time. Patch testing with updated test series and relevance assessment of positive reactions is a delicate task. Children, with lifelong use of MDs, risk contracting many allergies with potential cross-allergies. A question should be raised as to whether these low molecular weight acrylates should be used at all in devices constantly worn on the skin.
引用
- Ulriksdotter J, Svedman C, Bruze M, et al. Allergic contact dermatitis caused by dipropylene glycol diacrylate in the Omnipod® insulin pump. Br J Dermatol. 2021. PMID: 34510410. doi:10.1111/bjd.20751
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。