持续葡萄糖监测系统致儿童及成人接触性皮炎:调查与建议

Continuous glucose monitoring systems give contact dermatitis in children and adults despite efforts of providing less 'allergy- prone' devices: investigation and advice hampered by insufficient material for optimized patch test investigations.
PubMed 文献 · J Eur Acad Dermatol Venereol · 2021-01-01

该研究回顾了11例因疑似对胰岛素泵或葡萄糖传感器系统(特别是Dexcom G6®)发生过敏性接触性皮炎而转诊的1型糖尿病患者,分析了医疗器械提取物。多数患者为儿童,多数存在相关过敏,尤其是对丙烯酸异冰片酯过敏,该物质也存在于Dexcom G6®中。结论强调即使更换设备或暂时使用其他产品,已致敏患者仍可能发生过敏性接触性皮炎,并呼吁加强合作、完善斑贴试验系列及使用自身材料测试。

要点

  • 葡萄糖传感器系统(如Dexcom G6®)可致接触性皮炎,丙烯酸异冰片酯是重要过敏原。
  • 已致敏患者更换设备或产品后仍可能复发皮炎,需持续警惕。
  • 建议使用 adequate 斑贴试验系列并测试患者自身材料以明确诊断。

核心结论

  • 使用一种医疗器械致敏的患者,即使更换为另一种设备或暂时使用其他产品,仍可能发生过敏性接触性皮炎。

研究设计与样本

  • 回顾性分析11例因疑似对医疗器械(胰岛素泵或葡萄糖传感器系统,特别是Dexcom G6®)发生过敏性接触性皮炎而转诊的1型糖尿病患者,并分析器械提取物。

主要发现

  • 多数患者为儿童,多数存在相关过敏,尤其是对丙烯酸异冰片酯过敏,该物质也存在于Dexcom G6®中。

临床怎么用

  • 对疑似医疗器械相关接触性皮炎患者应进行斑贴试验,包括使用自身材料测试,并考虑丙烯酸异冰片酯等过敏原。

局限性

  • 医疗器械生产所用物质信息不统一且不完整,可供研究的材料稀缺,导致即使设备完善的实验室也难以充分调查。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Medical devices are increasingly being reported to cause contact allergic dermatitis reactions. OBJECTIVE: Review of patients with diabetes type I referred for suspected allergic contact dermatitis to insulin pump or glucose sensor systems. METHOD: We have reviewed 11 referred diabetes mellitus patients investigated for allergic contact dermatitis reactions to medical devices and specifically Dexcom G6® . Extracts from the medical devices were analysed. RESULTS: The majority of patients was children, the majority had relevant allergies and particularly allergy to isobornyl acrylate which was also found in the glucose sensor system Dexcom G6® . CONCLUSIONS: The following case reports bring in focus the fact that patients sensitized through use of one medical device and being advised the use of another, or find another product for a while useful, are not by necessity free from future episodes of allergic contact dermatitis. The case reports emphasize the need for collaboration since it is impossible for even well-equipped laboratories to properly investigate the medical devices when information on the substances used in production is not uniform and complete and material to investigate are scarce. The importance of adequate patch test series and testing with own material and furthermore the importance to re-analyse medical devices and re-analyse test data are emphasized.

来源:J Eur Acad Dermatol Venereol,查看原文。版权归原作者所有。

引用

  1. Svedman C, Bruze M, Antelmi A, et al. Continuous glucose monitoring systems give contact dermatitis in children and adults despite efforts of providing less 'allergy- prone' devices: investigation and advice hampered by insufficient material for optimized patch test investigations. J Eur Acad Dermatol Venereol. 2021. PMID: 33034101. doi:10.1111/jdv.16981