新型离体共聚焦激光扫描显微镜与H&E染色石蜡切片在基底细胞癌显微手术中的诊断准确性比较
研究评估了Histolog Scanner v1对148例基底细胞癌的544份新鲜标本生成离体共聚焦激光扫描显微镜(CLSM)图像并检测基底细胞癌组织的准确性。可分析525幅图像,灵敏度73%(95%CI=[65.27%; 80.47%]),特异度96%(95%CI=[93.40%; 97.60%]),钻孔活检中基底细胞癌检测灵敏度为100%。生成和评估CLSM图像的中位总时间为5.17分钟,最大20.17分钟,最小2.05分钟。
要点
- CLSM检测基底细胞癌的灵敏度为73%(95%CI=[65.27%; 80.47%]),特异度为96%(95%CI=[93.40%; 97.60%])。
- 钻孔活检中基底细胞癌检测灵敏度为100%。
- 生成和评估CLSM图像的中位总时间为5.17分钟(最大20.17分钟,最小2.05分钟)。
核心结论
- 共聚焦激光扫描显微镜是经典石蜡包埋或冷冻切片的省时且非常有效的替代方法,但所用显微镜的诊断准确性仍需提高。
研究设计与样本
- 研究纳入148例基底细胞癌的544份新鲜标本,经0.01%原黄素溶液染色后由Histolog Scanner v1生成数字显微图像,并将CLSM图像与对应H&E染色切片的组织学诊断进行比较。
主要发现
- 共可分析525幅图像,CLSM检测基底细胞癌的灵敏度为73%(95%CI=[65.27%; 80.47%]),特异度为96%(95%CI=[93.40%; 97.60%])。
- 钻孔活检中基底细胞癌检测灵敏度为100%。
- 生成和评估一幅CLSM图像的中位总时间为5.17分钟(最大20.17分钟,最小2.05分钟)。
- 最大挑战是压平标本以确保手术切缘的完整呈现。
临床怎么用
- CLSM是经典石蜡包埋或冷冻切片的省时且非常有效的替代方法,可因缩短住院时间或减少手术阶段间间隔而加快门诊治疗,从而改善患者治疗。
- 所用显微镜的诊断准确性仍需提高。
局限性
- 所用显微镜的诊断准确性仍需提高。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: For safe excision of malignant skin tumours, complete negative surgical margins are mandatory. The gold standard for analysis is frozen sections or paraffin-embedded haematoxylin and eosin (H&E)-stained slides. The production of H&E-stained slides is time-consuming (>20 h) while wounds remain unclosed. An upcoming method is confocal laser scanning microscopy (CLSM), a technique that scans unfixed fresh tissue rapidly. OBJECTIVE: Evaluation of the process to generate and analyse CLSM images and assessment of the accuracy to detect basal cell carcinoma (BCC) tissue. METHODS: Digital microscopic images were generated by the Histolog Scanner v1 from 544 fresh specimens of 148 BCCs that had been stained with a 0.01% proflavine solution. CLSM images were compared to the histological diagnoses of the corresponding H&E-stained slides. RESULTS: A total of 525 images could be analysed. The sensitivity was 73% (95% CI = [65.27%; 80.47%]), and the specificity was 96% (95% CI = [93.40%; 97.60%]). Detection of BCCs in punch biopsies was certainly detected (sensitivity of 100%). The median total time to generate and evaluate a CLSM image was 5.17 min (maximum 20.17 min and minimum 2.05 min). The greatest challenge was flattening the specimen to assure complete representation of the surgical margins. CONCLUSION: Confocal laser scanning microscopy is a time-saving and very effective alternative to classical paraffin-embedded or frozen sections. Patient treatment could be improved due to shorter hospital stays or faster outpatient therapy due to reduced intervals between surgical stages. Diagnostic accuracy of the microscope used still must be improved.