自动化计算机引导的银屑病面积与严重程度指数测量与受过培训医师比较的精确性和可重复性
该比较观察性研究在120例不同严重度的斑块状银屑病患者中,评估自动全身成像与计算机数字图像分析获得的自动化计算机引导PASI测量(ACPMs)与三位受过培训医师PASI测量的一致性(ICC 0·86)及重复测量可重复性(ICC 0·99),ACPMs在评估者内可靠性上优于三位医师(平均ICC 0·86)。
要点
- ACPMs与医师PASI计算一致性高(ICC 0·86,95%CI 0·80-0·90,平均绝对差2·5个PASI分)。
- 重复ACPMs可重复性极好(ICC 0·99,95%CI 0·98-0·99,平均绝对差0·5个PASI分)。
- ACPMs在评估者内可靠性上优于三位医师(平均ICC 0·86)。
核心结论
- 在120例银屑病患者中,ACPMs与受过培训医师相比精确性相似且可重复性更高。
研究设计与样本
- 比较观察性研究,纳入120例不同严重度的斑块状银屑病患者,采用自动全身成像和计算机数字图像分析获得ACPMs,并与三位受过培训医师的PASI测量比较。
主要发现
- 在120例完全可评估患者中,ACPMs与医师PASI计算的一致性高(ICC 0·86,95%CI 0·80-0·90,平均绝对差2·5个PASI分)。
- 重复ACPMs的可重复性显示极好ICC为0·99(95%CI 0·98-0·99),平均绝对差0·5个PASI分。
- ACPMs在评估者内可靠性上优于三位医师(平均ICC 0·86)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 局限性来自难以观察的身体部位以及无法在自动图像采集期间达到预定姿势的患者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The Psoriasis Area and Severity Index (PASI) represents the gold standard for psoriasis severity assessments but is limited by its subjectivity and low intra- and inter-rater consistency. OBJECTIVES: To investigate the precision and reproducibility of automated, computer-guided PASI measurements (ACPMs) in comparison with three trained physicians. METHODS: This was a comparative observational study assessing ACPMs attained by automated total-body imaging and computerized digital image analysis in a cohort of 120 patients affected by plaque psoriasis of various severities. The level of agreement between ACPMs and physicians' PASI measurements was calculated by the intraclass correlation coefficient (ICC). The reproducibility of ACPMs in comparison with physicians' PASI measurements was investigated by performing two successive 'repeat PASI calculations' in the same patients. RESULTS: The agreement between ACPMs and physicians' PASI calculations in 120 fully evaluable patients was high (ICC 0·86, 95% confidence interval 0·80-0·90, mean absolute difference 2·5 PASI points). Repeat ACPMs to measure the reproducibility showed an excellent ICC of 0·99 (95% confidence interval 0·98-0·99) with a mean absolute difference of 0·5 PASI points. The ACPMs thus outperformed the three physicians for intrarater reliability (mean ICC 0·86). CONCLUSIONS: The results of this first clinical study investigating ACPMs in 120 patients with psoriasis indicate a similar precision and higher reproducibility in comparison with trained physicians. Limitations arise from poorly observable body sites and from patients unable to attain predefined postures during automated image acquisition.
引用
- Fink C, Alt C, Uhlmann L, et al. Precision and reproducibility of automated computer-guided Psoriasis Area and Severity Index measurements in comparison with trained physicians. Br J Dermatol. 2018. PMID: 30218575. doi:10.1111/bjd.17200
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。