骨整合矩形块状种植体的生物力学特征:一项体内实验初步研究
本研究设计制造了12枚矩形块状种植体(RBI),植入3只成熟灵缇犬的下颌后牙区,每侧2枚,愈合12周后通过共振频率分析和扳手扭矩测试确认骨整合。随后取含种植体的骨块进行微型CT评估骨-种植体接触(BIC),并用万能测试系统进行拉出(拉伸)、颊向推(剪切)和远中向推(剪切)加载直至失败。结果显示骨整合RBI能承受模拟后牙区轴向、颊舌向和近远中向的生物力学挑战,界面剪切强度计算为14.4 MPa,超过人类最大咀嚼负荷的等效分力。
要点
- 骨整合RBI在骨量减少条件下可承受后牙区多方向生物力学负荷。
- 界面剪切强度计算为14.4 MPa,超过人类最大咀嚼负荷等效分力。
- 拉出、颊向和远中向力失败水平不同,骨失败模式包括小梁和皮质骨折。
核心结论
- 骨整合RBI能承受模拟后牙区轴向、颊舌向和近远中向的生物力学挑战,且在骨量减少条件下,其值超过人类最大咀嚼负荷的等效分力。
研究设计与样本
- 设计制造12枚矩形块状种植体(RBI),植入3只成熟灵缇犬的下颌后牙区,每侧2枚,愈合12周。
- 通过共振频率分析(RFA)和扳手扭矩测试确认骨整合,取含两枚RBI的骨块进行微型CT(μ-CT)评估骨-种植体接触(BIC),并用万能测试系统进行加载分析。
主要发现
- 拉出、颊向和远中向力失败发生在不同水平。
- 脱离后位点显示复杂的骨失败模式,包括小梁和皮质骨折,以及距BIC不同距离的剪切。
- 界面剪切强度计算为14.4 MPa。
关键数值与推荐
- 界面剪切强度计算值为14.4 MPa。
临床怎么用
- 骨整合RBI可能作为后牙区吸收牙槽嵴修复的一种解决方案,其生物力学性能在动物模型中超过人类最大咀嚼负荷等效分力。
局限性
- 本研究为初步体内实验,样本量小(3只犬、12枚RBI),结论需进一步临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: A novel implant design, the rectangular block implant (RBI), was investigated as a possible solution to the restoration of the posterior resorbed ridge. AIM: To maximally load test the osseo-integrated RBI in shear and tensile loads and relate these findings to known human masticatory loads as biomechanical proof of the study concept. MATERIALS AND METHODS: Twelve RBIs were design-manufactured and placed into posterior mandibular saddles in 3 mature greyhound dogs.-2 per left and right. After 12 weeks of healing, osseo-integration was confirmed using resonance frequency analysis (RFA) and wrench torque tests. Three bone blocks each with two RBIs were dissected and mounted in acrylic. Micro-computerized tomography (μ-CT) was performed to assess bone to implant contact (BIC), and load analysis was performed using a Universal Test System. Three force applications were conducted until failure: pull-out (tensile), buccal push from the lingual (shear) and distal push from the mesial (shear). The osteotomy sites were examined using light magnification and scanning electron microscopy (SEM). RESULTS: Pull-out, buccal and distal force failures occurred at differing levels. Post-detachment sites showed complex patterns of bone failure, including trabecular and cortical fracture, as well as shearing at varying distances from the BIC. Interfacial shear strength was calculated at 14.4 MPa. CONCLUSION: The osseo-integrated RBIs were able to withstand simulations of the demanding axially, bucco-lingually and mesio-distally oriented biomechanical challenges of the posterior saddle, under conditions of reduced bone volume. These values exceeded equivalent force components of maximal masticatory loads in humans.
引用
- Gazelakis E, Judge RB, Palamara JEA, et al. The biomechanical profile of an osseo-integrated rectangular block implant: A pilot in vivo experimental study. Clin Oral Implants Res. 2021. PMID: 34658084. doi:10.1111/clr.13834
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。