后牙种植单冠粘接与螺丝固位的24个月随机对照试验
结果显示两种固位方式在黏膜炎发生率、边缘骨吸收及修复体失败方面无显著差异,螺丝固位组出现1例螺丝松动,粘接组2个修复体检出粘接剂残留。
要点
- 后牙种植单冠粘接与螺丝固位的生物和技術并发症率无显著差异。
- 12个月黏膜炎发生率螺丝固位组14.2%、粘接固位组17.9%。
- 粘接组2个修复体(6.9%)基线时放射学检出粘接剂残留。
核心结论
- 整体锂二硅酸酯后牙种植单冠的固位方式对生物与技术创新并发症发生率无影响。
研究设计与样本
- 螺丝固位组将整体锂二硅酸酯修复体口外粘接到钛基底上;粘接固位组将整体锂二硅酸酯冠口内粘接到个性化钛基台上。
- 修复体在戴入2周内(基线)、12个月(n=46)和24个月(n=43)按改良FDI标准检查;基线和12个月通过标准化X线片评估骨丧失。
主要发现
- 12个月时黏膜炎(探诊出血阳性)发生率螺丝固位组为14.2%,粘接固位组为17.9%。
- 牙龈指数、菌斑指数及平均边缘骨丧失在0.03–0.15 mm之间,组间无显著差异。
- 粘接组基线时放射学检查在2个修复体(6.9%)检出粘接剂残留。
- 大多数病例(85.7%)实现了全数字化工作流程。
- 24个月时无修复体失败,无陶瓷崩裂;螺丝固位组1枚种植体发生螺丝松动。
临床怎么用
- 后牙种植单冠可依据临床情况选择粘接或螺丝固位,两者并发症风险相近。
- 粘接固位时需注意清除粘接剂残留,建议采用放射学检查确认。
- 螺丝固位需关注螺丝松动风险,定期复查咬合与邻接触点质量。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To compare the incidence of biological and technical complications of cemented and screw-retained monolithic lithium-disilicate implant-supported posterior single crowns. MATERIAL AND METHODS: Forty-one subjects with a total of 56 implants received randomly allocated 28 cemented and 28 screw-retained crowns. In the screw-retained group, monolithic lithium-disilicate restorations were luted to titanium bases extraorally. In the cemented group, monolithic lithium-disilicate crowns were cemented on individualized titanium abutments intraorally. All restorations were examined according to modified FDI criteria within 2 weeks of inserting the crowns (baseline) and after 12 (n = 46) and 24 (n = 43) months. Bone loss was evaluated by standardized radiographs at baseline and 12 months. RESULTS: After 12 months, the incidence of mucositis (positive bleeding on probing) was 14.2% (screw-retained) and 17.9% (cement-retained). The gingival and plaque index and a mean marginal bone loss between 0.03-0.15 mm showed no significant difference between the groups. In the cemented group, cement residues were detected at baseline at two restorations (6.9%) by radiographic examination. A complete digital workflow was realized in most cases (85.7%). At 24 months, no restoration had failed, and no chipping of the ceramic had occurred. In the screw-retained group, screw loosening occurred in one implant. In both groups, there was obvious deterioration in the quality of 32% of the occlusal and of 18% of the proximal contact points. CONCLUSIONS: The type of retention mode of monolithic implant-retained lithium-disilicate posterior crowns had no influence on the biological and technical complication rate.
引用
- Wolfart S, Rittich A, Groß K, et al. Cemented versus screw-retained posterior implant-supported single crowns: A 24-month randomized controlled clinical trial. Clin Oral Implants Res. 2021. PMID: 34547824. doi:10.1111/clr.13849
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。