全牙弓数字化种植印模的体外与体内准确性比较
研究比较全牙弓数字化种植印模在体外与体内条件下对固定修复的准确性。8例患者共10个无牙颌弓,每个制作含4个扫描体的钛杆,分别口内扫描8次(IOS组)和主模型扫描8次(MIOS组),以实验室扫描为参考。IOS组最大距离精度44±18µm、正确度32±19µm,MIOS组分别为31±16µm和30±14µm;口内条件中度影响精度和正确度,体外结果不能直接转化临床。
要点
- 口内条件对Trios 3口内扫描仪全牙弓种植印模的精度和正确度有中度影响。
- 体外模型扫描的准确性结果不能直接推广至临床口内扫描。
- 最大距离(1-4)的精度和正确度在体内外存在差异。
核心结论
- 口内条件中度影响Trios 3口内扫描仪全牙弓种植印模的精度和正确度。
- 体外准确性研究结果不能直接转化至临床领域。
研究设计与样本
- 自身对照研究,纳入8例患者(5女3男)共10个无牙颌弓,每弓含4-6枚骨结合种植体。
- 每个无牙颌弓制作含4个扫描体的实验性螺丝固位钛杆,分别进行口内扫描(IOS组,8次)和主模型扫描(MIOS组,8次),以实验室扫描为参考。
主要发现
- IOS组最大距离(1-4)的精度为44±18µm,正确度为32±19µm。
- MIOS组最大距离(1-4)的精度为31±16µm,正确度为30±14µm。
- IOS组最远扫描体间角度(1-4)的精度为0.22±0.14°,正确度为0.18±0.10°。
- MIOS组最远扫描体间角度(1-4)的精度为0.16±0.11°,正确度为0.07±0.05°。
临床怎么用
- 使用口内扫描仪进行全牙弓种植印模时,应考虑到口内环境对精度和正确度的影响。
- 不应将体外准确性研究的结果直接用于预测临床口内扫描的准确性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: The main objective of the study was to compare the accuracy of full-arch digital implant impressions for fixed dental prosthesis under in vitro and in vivo conditions. MATERIALS AND METHODS: Eight patients (five women and three men) with at least one edentulous arch and with 4-6 osseointegrated implants participated in this study. For each edentulous arch (n = 10), experimental screw-retained titanium bar with attached four scan bodies was fabricated. The bar containing four scan bodies was screw-retained intraorally on implants and scanned with Trios 3 intraoral scanner eight times (IOS group, in vivo). Then, the bar was attached to the master cast and scanned eight times again with the same intraoral scanner (MIOS group, in vitro). Finally, the bar with scan bodies was scanned 8 times with a laboratory scanner (reference). Precision and trueness were calculated for 3 distances and 3 angles between the scan bodies (1-2, 1-3, and 1-4) in IOS and MIOS groups. RESULTS: Precision and trueness for the largest distance (1-4) were found to be 44 ± 18 µm and 32 ± 19 µm for the IOS group and 31 ± 16 µm and 30 ± 14 µm for MIOS group, respectively. Precision and trueness for the angle between the most distant scan bodies (1-4) were 0.22 ± 0.14° and 0.18 ± 0.10° for the IOS group and 0.16 ± 0.11° and 0.07 ± 0.05° for MIOS group, respectively. CONCLUSIONS: Intraoral conditions moderately affected the precision and trueness of Trios 3 (3Shape) intraoral scanner. Results of in vitro accuracy studies cannot be directly transferred to the clinical field.
引用
- Rutkunas V, Gedrimiene A, Akulauskas M, et al. In vitro and in vivo accuracy of full-arch digital implant impressions. Clin Oral Implants Res. 2021. PMID: 34543478. doi:10.1111/clr.13844
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。