文献速递牙体牙髓病

R-Pilot与ProGlider在陡弯根管中的循环疲劳、表面形貌与扭转强度比较

Static Cyclic Fatigue Resistance in Abrupt Curvature, Surface Topography, and Torsional Strength of R-Pilot and ProGlider Glide Path Instruments.
PubMed 文献 · J Endod · 2021-09-15

本研究比较ProGlider与R-Pilot在人工陡弯根管中的循环疲劳抗力、ISO扭转抗力及临床使用后表面形貌变化。结果显示R-Pilot的循环疲劳与扭转抗力显著高于ProGlider,两者角 deflection 相似;使用后器械表面粗糙度未增加,但刃缘半径增大,ProGlider增幅更大。

要点

  • R-Pilot在陡弯根管中循环疲劳抗力显著高于ProGlider。
  • R-Pilot扭转抗力更高,但角 deflection 与ProGlider相似。
  • 临床使用后刃缘半径增大,ProGlider增幅大于R-Pilot。

核心结论

  • 在陡弯人工根管中,R-Pilot的循环疲劳强度高于ProGlider。
  • R-Pilot扭转抗力高于ProGlider,但角 deflection 值相似。
  • 陡弯根管预备未增加两种器械表面粗糙度,但导致刃缘半径增大。

研究设计与样本

  • 循环疲劳测试采用静态模型,人工根管具有90°弯曲角和2 mm弯曲半径。
  • 扭转抗力测试依据ISO 3630-1规范进行。
  • 使用光学轮廓仪和扫描电镜评估未使用与使用后器械表面形貌。
  • 基于微计算机断层扫描选取下颌第一磨牙近中根管进行 glide path 预备。

主要发现

  • R-Pilot的循环疲劳抗力和扭转抗力均显著高于ProGlider(P < .05)。
  • 两种器械的角 deflection 值相似(P < .05)。
  • ProGlider刃部区域的表面粗糙度值较大。
  • 未使用器械的切割刃粗糙度值显著大于使用后器械(P < .05)。
  • 使用后R-Pilot刃缘半径增加14%,ProGlider增加61%。

临床怎么用

  • 在陡弯根管 glide path 预备中,R-Pilot可能提供更高的循环疲劳与扭转抗力。
  • 临床使用后需关注器械刃缘半径变化,ProGlider的刃缘钝化更明显。

局限性

  • 研究为体外实验,使用人工根管模型,结果需结合临床验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: This study aimed to compare ProGlider (Dentsply Sirona, Ballaigues, Switzerland) and R-Pilot (VDW, Munich, Germany) instruments in terms of their cyclic fatigue resistance using an artificial stainless steel canal showing an abrupt apical curvature, torsional resistance according to the ISO specification, and topographic changes on the instrument surface after glide path management in mesial canals of mandibular first molars with the abrupt curvature selected based on their micro-computed tomographic examination. METHODS: Eighty instruments were used: 40 ProGlider (size 0.16, .02v taper) and 40 R-Pilot (size 0.125, .04 taper) instruments. The cyclic fatigue resistance was tested in a static test model using an artificial canal with an abrupt apical curvature (angle of curvature of 90° and radius of curvature of 2 mm). The torsional resistance test was performed according to ISO 3630-1 specifications. To determine surface topography of the unused and used instruments, mesial root canals of mandibular molars with an abrupt apical curvature were selected to prepare a glide path with either the ProGlider or R-Pilot instrument. An optical profilometer and scanning electron microscopy were used to determine the surface properties. Normally distributed torsional and cyclic resistance data were analyzed using the Student t test, whereas quantitative data obtained by the optical profilometer were analyzed with the Kruskal-Wallis H test with a 5% significance threshold. RESULTS: The R-Pilot showed significantly higher cyclic fatigue and torsional resistance than the ProGlider (P < .05). Angular deflection values were similar between instruments (P < .05). Measurements made from the blade area showed that the surface roughness values of the ProGlider were larger. Cutting blade measurements showed that unused instruments had significantly greater roughness values than used ones (P < .05). Although there was a 14% increase between the blade edge radii of the used and unused R-Pilot instruments, this difference was determined as 61% in ProGlider instruments. CONCLUSIONS: The R-Pilot exhibited greater cyclic fatigue strength than the ProGlider when tested in an artificial canal with an inner diameter of 1.0 mm and an abrupt apical curvature. Torsional resistance of the R-Pilot was higher than the ProGlider, but the angular deflection values were similar. Glide path preparation in a mesial root canal with an abrupt apical curvature did not increase the surface roughness of both instruments but resulted in a greater blade edge radius.

来源:J Endod,查看原文。版权归原作者所有。