文献速递口腔修复与种植

激光位移传感器分析即刻负载种植体稳定性变化并与共振频率分析灵敏度比较

Analysis of implant stability changes in immediate loading using a laser displacement sensor in vivo and comparison of its sensitivity with that of resonance frequency analysis.
PubMed 文献 · Clin Oral Implants Res · 2021-08-29

该研究在11只兔胫骨中植入即刻负载种植体,用激光位移传感器监测功能动态负载下的微动,并用共振频率分析(RFA)监测种植体稳定性商(ISQ)。结果显示微动在负载开始次日达峰值后逐渐平台化,ISQ变化与实际微动高度相关(r=-0.98,p<.01),但ISQ不能判定整合后的绝对稳定性。

要点

  • 即刻负载后种植体稳定性在负载开始后立即最低,随后逐渐增高。
  • ISQ变化与实际微动高度相关(r=-0.98,p<.01),可用于监测稳定性变化。
  • ISQ不能判定种植体整合后的绝对稳定性幅度。

核心结论

  • 即刻负载种植体在负载开始后稳定性最低,之后逐渐增高。
  • RFA可用于检查稳定性变化和初始稳定性,但不能确定整合后稳定性的绝对幅度。

研究设计与样本

  • 在11只兔胫骨中植入即刻负载种植体,用激光位移传感器监测微动。
  • 功能动态负载每周5天、持续6周,并用RFA监测ISQ。

主要发现

  • 几乎全部负载种植体的微动在负载开始次日增至峰值,随后逐渐达到平台期。
  • 负载种植体的ISQ变化与实际微动变化高度相关(r=-0.98,p<.01)。
  • 初始固定时ISQ值与测得微动相关(r=0.73,p<.05),但与获得整合的种植体微动不相关。
  • ISQ与组织形态计量数据之间未见密切相关性。

临床怎么用

  • RFA可用于监测即刻负载种植体的稳定性变化和评估初始稳定性。
  • 不宜仅凭ISQ判定种植体整合后的绝对稳定性幅度。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The aim of this study was to analyze the stability changes in immediately loaded implants by using an in vivo quantitative measurement of micromotion under functional dynamic loading and to verify the sensitivity of Resonance Frequency Analysis (RFA) as compared to that of actual micromotion. MATERIALS AND METHODS: The micromotions of immediately loaded implants placed in the tibia of 11 rabbits were monitored using a laser displacement sensor. Functional dynamic loading forces were applied 5 days a week for 6 weeks. The implant stability quotient (ISQ) was monitored using RFA. RESULTS: The micromotion of the almost-loaded implants increased to peak values the day after loading was started and subsequently reached a plateau gradually. The ISQ changes in the loaded implants closely correlated with the alterations of the actual micromotion (r = -0.98, p < .01). Although the ISQ value itself correlated with the measured micromotion at the time of initial fixation (r = 0.73, p < .05), it did not correlate with the micromotion of the implant that acquired integration. No close correlation was observed between the ISQ and the histomorphometrical data. CONCLUSION: The immediately loaded implants showed the lowest stability immediately after the start of loading, which gradually increased thereafter. RFA is considered a useful method for examining stability changes and initial stability; however, it cannot determine the absolute magnitude of the stability after integration.