体素大小与滤镜对CBCT检测MB2根管的影响
本研究评估体素大小和滤镜应用对锥形束CT检测第二近中颊根管(MB2)的影响。采用OP300 CBCT系统以80、125、200微米三种体素协议扫描40颗第一磨牙,由5名口腔放射科医生在三种滤镜条件下评估。结果显示80微米体素组检测准确性更高,而增强滤镜未产生显著影响。
要点
- 80微米体素大小可提高MB2根管检测的准确性
- 增强滤镜应用对MB2检测准确性无显著影响
- 观察者内一致性中位值为0.70,观察者间为0.56
核心结论
- 较小体素尺寸提高了检测MB2根管的准确性,而增强滤镜未提高准确性。
研究设计与样本
- 采用OP300 CBCT系统以80、125、200微米三种体素协议扫描40颗第一磨牙(20颗有MB2根管,20颗无),由5名口腔放射科医生在无滤镜、sharpen 1×滤镜和sharpen 2×滤镜三种条件下评估。
主要发现
- 80微米体素联合sharpen 1×滤镜图像组的敏感性、准确性和阴性预测值最高。
- 80微米体素无滤镜组的特异性和阳性预测值最高。
- 80微米组的受试者工作特征曲线下面积高于125微米和200微米组(P < .05)。
- 滤镜使用(P = .22)及滤镜与体素大小的交互作用(P = .88)均无显著差异。
- 观察者内和观察者间一致性的中位值分别为0.70和0.56。
临床怎么用
- 在临床检测MB2根管时,优先选择较小体素尺寸(如80微米)的CBCT扫描方案。
- 增强滤镜的应用可能不会带来额外益处,可根据需要选择是否使用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study assessed the influence of voxel size and filter application in detecting second mesiobuccal (MB2) canals in cone-beam computed tomographic (CBCT) images. METHODS: Using the OP300 CBCT system (Instrumentarium, Tuusula, Finland) and 3 voxel size protocols (80 μm, 125 μm, and 200 μm), we scanned 40 first molars: 20 with an MB2 canal and 20 without. All molars received silver palladium pins on the palatal root, whereas the non-MB2 molars were also filled with gutta-percha. Five oral radiologists assessed the presence of an MB2 canal under 3 filter application conditions: without filter, with sharpen 1 × filter, and with sharpen 2 × filter. Intra- and interobserver reproducibility was evaluated using the weighted kappa index. We compared the area under the receiver operating characteristic curves with SPSS Statistics v.20.0 (IBM Corp, Armonk, NY) using 2-way analysis of variance and the Tukey post hoc test with 5% significance level. RESULTS: Our analysis found median intra- and interobserver agreement values of 0.70 and 0.56, respectively. The 80-μm voxel with sharpen 1 × filter image group had the highest sensitivity, accuracy, and negative predictive values. As for specificity and positive predictive, the 80-μm voxel group without filter application presented the highest values. The areas under the receiver operating characteristic curve were higher in the 80-μm groups than in the 125-μm and 200-μm voxel size groups (P < .05). We found no differences among the filters used (P = .22) or for the filter-voxel size interactions (P = .88). CONCLUSIONS: A smaller voxel size increased the accuracy in detecting MB2 canals, whereas the enhancement filters did not.
引用
- Mouzinho-Machado S, Rosado LPL, Coelho-Silva F, et al. Influence of Voxel Size and Filter Application in Detecting Second Mesiobuccal Canals in Cone-beam Computed Tomographic Images. J Endod. 2021. PMID: 34166686. doi:10.1016/j.joen.2021.06.011
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。