七种根管冲洗方法在开放根尖上颌中切牙中产生的根尖压力:一项实验室研究
该实验室研究在3D打印的开放根尖(最大直径2.1mm)上颌中切牙上,测试七种根管冲洗方法产生的根尖压力。以5.73mmHg为临界阈值,EndoVac、SAF、XP-endo Finisher和UAI在任何深度或强度下均未超过阈值;Er:YAG激光仅在偏离厂商推荐的频率设置下超标;EDDY在TL及RinsEndo在任何深度下100%超标。
要点
- EndoVac、SAF、XP-endo Finisher和UAI在开放根尖下根尖压力均未超过5.73mmHg临界阈值。
- EDDY在牙长(TL)处及RinsEndo在任何深度下100%测量超过临界阈值。
- Er:YAG激光仅在偏离厂商推荐的频率设置时超过临界阈值。
核心结论
- 在模拟开放根尖的上颌中切牙中,EndoVac、Er:YAG激光激活、UAI、SAF和XP-endo Finisher产生的根尖压力低于5.73mmHg临界阈值,而EDDY和RinsEndo产生的根尖压力超过临界阈值。
研究设计与样本
- 实验室研究,在3D打印的开放根尖(最大直径2.1mm)上颌中切牙上测试七种根管冲洗方法,每种实验重复十次,采用描述性分析(最大值、均值、标准差、95%CI)。
主要发现
- EndoVac、SAF、XP-endo Finisher和UAI在任何插入深度或强度水平下均未超过临界阈值。
- Er:YAG激光激活仅在偏离厂商推荐的频率设置下超过临界阈值。
- EDDY在牙长(TL)处及RinsEndo在任何插入深度下100%的测量超过临界阈值。
- EDDY在TL-1、TL-2和TL-3 mm处分别有30%、10%和20%的测量超过临界阈值。
关键数值与推荐
- 根尖冲洗液挤出的临界阈值设定为5.73mmHg(中心静脉压下限:5.88 ± 0.15mmHg)。
- 压力传感器量程为0至120mmHg,响应时间≤0.5ms。
临床怎么用
- 在开放根尖病例中,优先选择EndoVac、SAF、XP-endo Finisher或UAI以降低根尖压力风险。
- 使用EDDY或RinsEndo时需谨慎,因其在特定条件下根尖压力可能超过临界阈值。
- Er:YAG激光激活应严格遵循厂商推荐的频率设置以避免根尖压力超标。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIM: A laboratory study to determine the apical pressure generated by seven canal irrigation methods in an anterior tooth with an open apex. METHODOLOGY: Canal irrigation was performed on a 3D-printed central maxillary incisor with an open apex (maximum diameter of 2.1 mm). Ultrasonically activated irrigation (UAI), sonic activation (EDDY), negative pressure irrigation (EndoVac), the self-adjusting file (SAF) and the XP-endo Finisher were employed at tooth length (TL), TL-1 mm, TL-2 mm and TL-3 mm. UAI was tested at three intensity levels additionally. Hydrodynamic irrigation with RinsEndo was performed in the pulp chamber, at the canal orifice, the coronal third, the middle of the canal and at TL. Er:YAG laser activation, at four frequency settings, was performed in the pulp chamber and at the orifice of the canal. The pressure of the fluid towards the canal terminus generated by activation was directly transferred to a pressure sensor with a range of 0 to 120 mmHg and a response time of ≤0.5 ms. The critical threshold for apical extrusion of the irrigant was set at 5.73 mmHg (lower limit of the central venous pressure: 5.88 ± 0.15 mmHg). Each experiment was repeated ten times. The tests were followed by descriptive analyses (maximum, mean, standard deviation, 95% confidence interval). RESULTS: EndoVac, the SAF, the XP-endo Finisher, and UAI never exceeded the critical threshold at any insertion depth or intensity level. Er:YAG laser activation exceeded the critical threshold exclusively at frequency settings that varied from the manufacturer's recommendation. EDDY at TL and RinsEndo at any insertion depth exceeded the critical threshold in 100% of the measurements. EDDY at TL-1,-2, and-3 mm crossed the critical threshold in 30%, 10%, and 20% of the measurements, respectively. CONCLUSIONS: In a simulated maxillary central incisor with an open apex, irrigation with EndoVac, Er:YAG laser activation, UAI, the SAF, and the XP-endo Finisher generated apical pressures below the critical threshold of 5.73 mmHg. By contrast, using EDDY and RinsEndo for irrigation produced higher apical pressures that exceeded the critical threshold.
引用
- Magni E, Jäggi M, Eggmann F, et al. Apical pressures generated by several canal irrigation methods: A laboratory study in a maxillary central incisor with an open apex. Int Endod J. 2021. PMID: 34022070. doi:10.1111/iej.13575
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。