三种电子根尖定位仪准确性的实验室研究:包埋介质与影像评估的影响

A laboratory study of the accuracy of three electronic apex locators: influence of embedding media and radiographic assessment of the electronic apical limit.
PubMed 文献 · Int Endod J · 2021-03-30

结果显示导电凝胶使多数读数过度延伸,三种设备在0.0/APEX标记配藻酸盐时准确性相似,影像视角无差异。

要点

  • 导电凝胶较藻酸盐显著增加电子读数过度延伸的风险。
  • 三种设备在0.0/APEX标记配藻酸盐时定位根尖孔准确性相似。
  • 下颌前磨牙藻酸盐包埋下,颊舌与近远中影像视角评估无差异。

核心结论

  • 与藻酸盐相比,ProTrain导电凝胶对多数电子读数产生负面影响。
  • 三种定位仪在0.0/APEX标记配藻酸盐时定位根尖孔准确性相似。
  • 藻酸盐包埋下颌前磨牙时,两种放射影像视角评估0.5标记无差异。

研究设计与样本

  • 比较藻酸盐与导电凝胶两种包埋介质及三种电子根尖定位仪。
  • 采用双因素方差分析、Tukey检验和卡方检验,显著性水平5%。

主要发现

  • 使用ProTrain导电凝胶时,除Root ZX II在0.5标记外,过度延伸读数发生率显著更高(P < 0.05)。
  • 以0.0/APEX标记定位根尖孔时,Root ZX、CanalPro和Apex ID准确性相似。
  • 放射影像法评估中,设备间及影像视角间未发现差异。

关键数值与推荐

  • 以放大10倍下10号K锉达根尖大孔的长度为实际长度(AL)。
  • 在三种定位仪0.5标记读数后拍摄颊舌向与近远中向两张根尖片。
  • 计算电子读数与AL的差值及影像上锉尖至根尖的距离。

临床怎么用

  • 实验室条件下优先使用藻酸盐作为包埋介质以避免读数过度延伸。
  • 三种定位仪在0.0/APEX标记配藻酸盐时可互换用于根尖孔定位。
  • 下颌前磨牙藻酸盐包埋时影像视角选择不影响根尖止点评估。

局限性

  • 离体牙实验室研究,结果外推至临床需谨慎。
  • 仅纳入单根下颌前磨牙,未涉及其他牙位或根管形态。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: To evaluate several factors that might interfere with the use of electronic root canal length measurement devices (ERCLMDs) in a laboratory setting, including two different embedding media (alginate and electroconductive gel), three different types of devices and the radiographic view on the assessment of the electronic readings. METHODOLOGY: Thirty single-rooted extracted human mandibular premolars were selected. After access and canal pre-flaring, a size 10 K-file was inserted in the canal up to the major apical foramen under magnification (×10), and this length was recorded as the actual length (AL) of the canal. Teeth were mounted on a specific endodontic training kit platform (ProTrain). Two different embedding media (electroconductive gel and alginate) and three different ERCLMDs: Apex ID (SybronEndo, Glendora, CA, USA), CanalPro Apex Locator (Coltene-Endo, Cuyahoga Falls, OH, USA) and the Root ZX II (J. Morita, Tokyo, Japan), were tested. Following the measurement at the 0.5 mark of each ERCLMD using alginate, two periapical radiographs (bucco-lingual and proximal views) were taken. The difference between the electronic readings and the AL, as well as the distance from the file tip to the apex of the roots on the radiographs, was calculated. Data were analysed statistically (two-way anova, Tukey's and chi-squared tests) at 5% significance level. RESULTS: The incidence of over-extended readings was significantly greater (P < 0.05) when using the ProTrain electroconductive gel, except for Root ZX II at the 0.5 mark. Root ZX, CanalPro and Apex ID had similar accuracy when the 0.0/APEX mark was used to locate the foramen. For the radiographic method, no difference was found between the devices or radiographic views. CONCLUSION: Compared with alginate, the conductive gel of the ProTrain kit negatively affected most of the electronic readings. The three evaluated ERCLMDs had similar accuracy in locating the apical foramen when using the mark 0.0/APEX and alginate as the embedding media. Although the ProTrain platform allows radiographs to be exposed in both bucco-lingual and proximal views, no difference was found between the views when evaluating the apical limit provided by the 0.5 mark in mandibular premolars embedded in alginate.

来源:Int Endod J,查看原文。版权归原作者所有。

引用

  1. Iparraguirre Nuñovero MF, Piasecki L, Segato AVK, et al. A laboratory study of the accuracy of three electronic apex locators: influence of embedding media and radiographic assessment of the electronic apical limit. Int Endod J. 2021. PMID: 33683733. doi:10.1111/iej.13507

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。