10种锥形束CT设备对牙髓细微结构检测能力的评估
本研究使用含牙颌面体模,在无金属和含金属桩两种条件下扫描10种CBCT设备,由3名观察者评估狭窄根管、峡部和根尖三角洲分支的显示能力。结果显示10种设备对牙髓任务表现存在差异(P < .05),3种设备表现更优,但含金属时仅2种设备能保持高水平检测。
要点
- 不同CBCT设备对牙髓细微结构的检测能力存在显著差异
- 含金属桩时多数设备检测能力下降,仅2种设备保持高水平
- 选择CBCT设备时应考虑其牙髓细微结构显示性能
核心结论
- 评估的牙髓任务依赖于CBCT设备,且其检测受金属存在的影响。
研究设计与样本
- 使用专用含牙颌面体模,在无金属和含金属桩两种条件下,用10种CBCT设备各扫描2次,以工业微CT图像为参考进行配准。
- 3名经验丰富的观察者采用1(最佳)至10(最差)的分类等级评估狭窄根管、峡部和根尖三角洲分支的显示能力。
- 使用Fleiss kappa统计量计算每种CBCT设备的观察者内和观察者间一致性,并基于观察者评分应用一般线性混合模型比较图像质量(α = .05)。
主要发现
- 10种CBCT设备在评估的牙髓任务上表现不同(P < .05),其中3种设备在牙髓特征检测方面表现更好。
- 在存在金属的情况下,仅2种设备能够保持高水平的牙髓特征检测。
临床怎么用
- 临床选择CBCT设备用于牙髓细微结构评估时,应参考其对狭窄根管、峡部和根尖三角洲分支的检测性能。
- 对于含金属桩的病例,需注意多数CBCT设备检测能力可能下降,应优先选择在含金属条件下仍保持高检测水平的设备。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The purpose of this study was to classify 10 cone-beam computed tomographic (CBCT) devices using a ranking model according to the detection of fine endodontic structures. METHODS: A dedicated dentate anthropomorphic phantom was scanned 2 times using 10 CBCT devices without any metal (metal-free condition) and with an endodontically treated tooth containing a metallic post (metal condition). A reference image acquired on an industrial micro-CT scanner was used to register all CBCT images, yielding corresponding anatomic slices. Afterward, 3 experienced observers assessed all acquired CBCT images for their ability to assess a narrow canal, isthmus, and apical delta ramification using a categoric rank from 1 (best) to 10 (worst). Fleiss kappa statistics were used to calculate intra- and interobserver agreements for each CBCT device separately. Based on the observers' scores, general linear mixed models were applied to compare image quality among different CBCT devices for performing endodontic diagnostic tasks (α = .05). RESULTS: The 10 CBCT devices performed differently for the evaluated endodontic tasks (P < .05), with 3 devices performing better for endodontic feature detection. Yet, in the presence of metal, only 2 devices were able to keep a high level of endodontic feature detection. CONCLUSIONS: The evaluated endodontic tasks were CBCT device dependent, and their detection was influenced by the presence of metal.
引用
- Pinto JC, Wanderley VA, de Faria Vasconcelos K, et al. Evaluation of 10 Cone-beam Computed Tomographic Devices for Endodontic Assessment of Fine Anatomic Structures. J Endod. 2021. PMID: 33691170. doi:10.1016/j.joen.2021.02.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。