新型活塞运动反角手机用于根管预备的研发
本研究评估新型活塞运动反角手机的最佳振幅与重量,并在树脂根管块上比较其与手用K锉及镍钛往复锉的预备效果。结果显示最佳条件为振幅1.35 mm、重量61.0 g,活塞组在根尖孔2.0-3.0 mm处中心偏移小于手动组,与镍钛组无显著差异。
要点
- 最佳参数为振幅1.35 mm、重量61.0 g。
- 活塞组中心定位能力与镍钛往复组相当,优于手动组。
- 活塞组预备时间与镍钛组无显著差异。
核心结论
- 新型反角手机可实现高效根管预备和探查,并可能避免台阶形成、器械分离和穿孔等根管意外。
研究设计与样本
- 设计原型反角手机,由20名操作者分3组在树脂根管块上完成预备:活塞组使用不锈钢#25手用H锉,手动组使用不锈钢#15-25 K锉,镍钛组使用Reciproc Blue R25镍钛往复锉配合根管马达。
- 测量根管中心线偏移量和预备所需时间并进行统计分析。
主要发现
- 最佳条件为振幅1.35 mm、重量61.0 g。
- 所有组均观察到根管中心线偏移。
- 距根尖孔2.0-3.0 mm处,活塞组或镍钛组与手动组之间差异有统计学意义,活塞组与镍钛组之间无显著差异。
- 镍钛组和活塞组中心偏移量最小;活塞组使用#25 H锉表现出良好的中心定位能力,与镍钛锉相似,且预备迅速。
关键数值与推荐
- 活塞组或镍钛组与手动组之间预备时间差异有统计学意义,活塞组与镍钛组之间无显著差异(P < .05)。
临床怎么用
- 新型反角手机使用#25 H锉时中心定位能力与镍钛往复锉相似,且预备速度相当,可作为根管预备的潜在选择。
局限性
- 研究在树脂根管块上进行,非真实牙根管,结论需临床进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The purpose of this study was to assess the optimal amplitude and weight of the newly developed contra-angle handpiece. The handpiece uses piston movement without using an endodontic motor and enables a safe, quick, and reliable canal preparation. METHODS: A prototype handpiece was designed. Instrumentation was performed on root canal resin blocks by 20 operators in 3 groups: the prototype handpiece with an H file (a stainless steel #25 manual H file, the piston group), a manually standardized technique with a K file (stainless steel #15-25 K files, the manual group), and a nickel-titanium (NiTi) reciprocating file with an endodontic motor (Reciproc Blue R25 [VDW, Munich, Germany], the NiTi group). Transportation of the canal center line and the time required for preparation were measured and statistically analyzed. RESULTS: The optimal condition was an amplitude of 1.35 mm and a weight of 61.0 g. Transportation of the canal center was observed in all groups. A statistically significant difference was found at 2.0-3.0 mm from the apical foramen between the piston or NiTi group and the manual group, but no significant difference was found between the piston and NiTi groups. The least transportation was found in the NiTi and piston groups. The handpiece with a #25 H file demonstrated a good centering ability, similar to the NiTi file, which enabled speedy preparation. The time required for preparation between the piston or NiTi group and the manual group was statistically different. No significant difference was observed between the piston and NiTi groups (P < .05). CONCLUSIONS: We concluded that the newly designed handpiece achieved efficient canal preparation and negotiation. The handpiece could avoid endodontic accidents, including ledge formation, instrument separation, and perforation.
引用
- Noguchi I, Suzuki K, Yoshida T, et al. Development of Innovative Contra-angle Handpiece Device with Piston Movement for Root Canal Preparation. J Endod. 2021. PMID: 33434564. doi:10.1016/j.joen.2020.12.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。