两种种植体颈部设计对25年种植体周愈合与功能骨结合影响的前瞻性随机临床试验

Prospective randomized clinical trial evaluating the effects of two different implant collar designs on peri-implant healing and functional osseointegration after 25 years.
PubMed 文献 · Clin Oral Implants Res · 2021-01-25

本研究比较了两种机加工颈部长度和设计的种植体对种植体周愈合的影响。种植体存活率为99%。

要点

  • 较长机加工颈部种植体25年边缘骨丧失显著大于较短颈部种植体。
  • 内部连接微织构种植体角化组织高度大于外部连接微织构种植体(0.74mm vs 0.51mm,p=0.01)。
  • 两组在菌斑、探诊深度、出血和溢脓方面无显著差异,种植体存活率为99%。

核心结论

  • 较短机加工颈部(1.2mm)种植体将骨丧失限制在1mm,而较长机加工颈部种植体在微间隙位于嵴顶时25年功能后骨丧失超过1.5mm。
  • 内部连接设计和种植体表面微织构未带来更好的骨保存效果。

研究设计与样本

  • 前瞻性随机临床试验,比较三种种植体:微织构表面+3.6mm长内部连接机加工颈部、微织构表面+1.2mm短外部连接机加工颈部、机加工表面+1.2mm短外部连接机加工颈部。

主要发现

  • 较长机加工颈部微织构种植体平均边缘骨丧失为-1.77mm ± 0.18,显著大于机加工短颈部种植体(-0.85mm ± 0.18,p < .001)和微织构短颈部种植体(-1.00 ± 0.18mm,p < .001)。
  • 内部连接微织构种植体角化组织高度为0.74mm ± 0.10,大于外部连接微织构种植体的0.51 ± 0.08(p = 0.01)。
  • 菌斑(p = 0.78)、探诊深度(p = 0.42)、出血(p = 0.07)和溢脓(p = 1.00)无显著差异。
  • 种植体存活率为99%。

关键数值与推荐

  • 较短机加工颈部(1.2mm)种植体骨丧失限制在1mm以内。
  • 较长机加工颈部(3.6mm)种植体25年功能后骨丧失超过1.5mm。
  • 内部连接微织构种植体角化组织高度0.74mm ± 0.10,外部连接微织构种植体0.51 ± 0.08(p = 0.01)。
  • 种植体存活率99%。

临床怎么用

  • 选择较短机加工颈部(1.2mm)种植体可能有助于减少长期边缘骨丧失。
  • 内部连接设计可能有利于维持角化组织高度。
  • 需长期监测种植体周骨水平,尤其对于较长机加工颈部设计。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Evaluate the effects of two different machined-collar lengths and designs on peri-implant healing. MATERIAL AND METHODS: An implant with a microtextured surface and 3.6mm-long internal-connection machined collar was compared to two implants that had an identical 1.2mm-long external-connection machined collar, but one had the microtextured surface while the other's was machined. Participants received the three implants, with microgap at the crest, alternately at five sites between mental foramen, and a full-arch prosthesis. Peri-implant bone levels were measured after 23 to 26 years of function. Keratinized tissue height, plaque, probing depth, bleeding, and purulence were also evaluated. Descriptive and mixed models for repeated\measures analyses were used, with Bonferroni correction for pairwise comparisons. RESULTS: Twenty-two participants (110 implants) were evaluated at the 25-year examination. Microtextured implants with the longer machined collar had significantly greater mean marginal bone loss (-1.77mm ± 0.18, mean ± SE) than machined (-0.85mm ± 0.18, p < .001) and microtextured (-1.00 ± 0.18mm, p < .001) implants with the shorter machined collar. Keratinized tissue height was greater for internal-connection (0.74mm ± 0.10) versus external-connection (0.51 ± 0.08, p = 0.01) microtextured implants. No differences were observed for plaque (p = 0.78), probing depth (p = 0.42), bleeding (p = 0.07), and purulence (p = 1.00). Implant survival rate was 99%. CONCLUSIONS: Implants with the 1.2mm machined collar limited bone loss to 1mm, while those with the longer machined collar showed > 1.5mm loss after 25 years of function with microgap at the crest. Internal-connection design and fixture surface microtexturing did not result in greater bone preservation. ClinicalTrials.gov Identifier: NCT03862482.