连续旋转与最优滑行路径运动对滑行路径器械扭矩、拧入力和成形能力的比较
结果显示OGP运动在22mm预备时产生更低的顺时针扭矩和拧入力,所有#10/0.05器械均断裂,各组间根管偏移无显著差异。
要点
- OGP运动在22mm预备时顺时针扭矩和拧入力低于连续旋转。
- 所有#10/0.05器械在实验中均发生断裂。
- 各组间根管偏移值无显著差异。
核心结论
- 与连续旋转相比,OGP运动产生更低的拧入力和顺时针扭矩,且根管偏移相似。
主要发现
- 所有#10/0.05器械均断裂。
- 在22mm预备中,OGP运动的顺时针扭矩和拧入力低于连续旋转(P < .05)。
- 在20mm预备中,#15/0.03器械OGP运动的拧入力低于连续旋转(P < .05)。
- 连续旋转在22mm预备中的顺时针扭矩和拧入力高于20mm预备(P < .05),而OGP运动两阶段无显著差异。
- 各组间根管偏移值无显著差异。
临床怎么用
- 在双弯根管预备中,OGP运动可降低拧入力和顺时针扭矩,可能减少器械 stress。
- #15/0.03 HyFlex EDM与#13/0.04原型MANI器械性能相似。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study aimed to analyze torque/force generation and transportation in double-curved canals instrumented with 3 types of glide path files using optimum glide path (OGP) motion in comparison with continuous rotation. METHODS: Sixty simulated double-curved canals were prepared with #10/0.05 or #15/0.03 HyFlex EDM Glidepath files (Coltene/Whaledent, Altstätten, Switzerland) or a #13/0.04 prototype MANI Glidepath file (Tochigi, Japan) using OGP motion or continuous rotation (n = 10 each). Canals were sequentially prepared to 20 mm and 22 mm (full working length) using automated root canal instrumentation and a torque/force analyzing device. Transportation was calculated at 1-9 mm from the apex. Data were compared using 2-way analysis of variance followed by a post hoc simple main effect test with Bonferroni correction and a Kruskal-Wallis test (α = 5%). RESULTS: All #10/0.05 instruments fractured. In the 22-mm preparation, the OGP motion resulted in lower clockwise torque and screw-in force than did continuous rotation (P < .05). In the 20-mm preparation, #15/0.03 instruments recorded a lower screw-in force for OGP motion than for continuous rotation (P < .05). Comparing the 2 preparation phases, OGP motion generated no significant differences; however, continuous rotation developed higher clockwise torque and screw-in force in the 22-mm preparation than in the 20-mm preparation (P < .05). There was no significant difference among the tested groups for transportation values. CONCLUSIONS: Compared with continuous rotation, OGP motion generated less screw-in force, lower clockwise torque, and similar transportation. The #15/0.03 HyFlex EDM instrument and the #13/0.04 prototype MANI instrument performed similarly well.
引用
- Htun PH, Ebihara A, Maki K, et al. Comparison of Torque, Screw-in Force, and Shaping Ability of Glide Path Instruments in Continuous Rotation and Optimum Glide Path Motion. J Endod. 2020. PMID: 33002538. doi:10.1016/j.joen.2020.09.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。