相邻牙金属桩对锥形束CT检测垂直根折的影响
本研究评估相邻牙含金属桩及采集参数对锥形束CT检测垂直根折的影响。对10颗单根牙在模拟根折前后进行成像,设置不同相邻牙条件、管电流和MAR算法,由5名口腔颌面放射科医师评估。结果显示相邻双牙均有金属桩时检测能力下降,提高管电流至8 mA可改善,MAR工具效果不佳。
要点
- 相邻双牙均有金属桩时,锥形束CT检测垂直根折的Az值显著降低。
- 管电流从4 mA提高至8 mA可显著提高垂直根折检测的Az值。
- 金属伪影减少(MAR)算法在此类情况下似乎无效。
核心结论
- 相邻双牙均有金属桩修复时,锥形束CT检测垂直根折的能力受损。
- 将管电流提高至8 mA可能有助于该诊断任务。
- MAR工具在此类情况下似乎无效。
研究设计与样本
- 对10颗单根牙在模拟垂直根折前后进行锥形束CT成像。
- 设置不同相邻牙条件(对照、1颗相邻牙、双侧相邻牙)、管电流(4、8、10 mA)和MAR(无/有)。
- 由5名口腔颌面放射科医师采用5分法评估垂直根折存在与否。
- 采用双因素方差分析比较诊断值,显著性水平为5%。
主要发现
- 垂直根折检测的Az值受相邻牙存在和毫安变化影响。
- 在4 mA下,双侧修复牙存在时Az值显著低于对照组(P ≤ .05)。
- 双侧修复牙存在时,8 mA的Az值显著高于4 mA(P ≤ .05)。
临床怎么用
- 对于相邻牙有金属桩的病例,可考虑将管电流提高至8 mA以改善垂直根折检测。
- 不应依赖MAR工具来克服金属桩伪影对垂直根折检测的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The aim of the present study was to evaluate the detection of vertical root fracture (VRF) in the presence of adjacent teeth restored with a metal post and the influence of acquisition parameters (tube current and metal artifact reduction [MAR] algorithm) on this diagnostic task. METHODS: Cone-beam computed tomographic images of 10 single-rooted teeth were acquired before and after the simulation of VRF. The acquisitions were set up to simulate different conditions regarding the presence of adjacent teeth restored with a metal post (control, 1 adjacent tooth, and both adjacent teeth), different tube currents (4, 8, and 10 mA), and the use of MAR (without MAR and with MAR). Images were assessed by 5 oral and maxillofacial radiologists using a 5-point scale for the presence of VRF. Diagnostic values were calculated and compared by 2-way analysis of variance (significance level of 5%). RESULTS: The area under the receiver operating characteristic curve (Az) values for VRF detection were affected by the presence of adjacent teeth and the variation of milliamperes. For 4 mA, when both restored teeth were present, Az values were significantly lower than the control group (P ≤ .05). In the presence of both restored teeth, 8 mA presented significantly higher Az values compared with 4 mA (P ≤ .05). CONCLUSIONS: The presence of both adjacent teeth restored with a metal post impairs VRF detection; however, an increase in tube current up to 8 mA may aid in this diagnostic task. Moreover, the MAR tool does not seem to be efficient in those cases.
引用
- Gaêta-Araujo H, de Oliveira Reis L, Nascimento EHL, et al. Influence of Metal Post in Adjacent Teeth in the Detection of Vertical Root Fracture Using Cone-beam Computed Tomography with Different Acquisition Parameters. J Endod. 2020. PMID: 32835757. doi:10.1016/j.joen.2020.08.013