牙科磁共振成像评估上颌磨牙根分叉病变的体内准确性
这项前瞻性研究纳入22例重度牙周炎患者,对192个根分叉入口用牙科磁共振成像(dMRI)与锥形束CT(CBCT)评估水平根分叉病变(FI)程度。dMRI与CBCT一致性高(偏倚0.17 mm),检测FI的敏感度和特异度分别为98%和99%,提示dMRI评估上颌磨牙FI准确可靠。
要点
- dMRI检测上颌磨牙根分叉病变的敏感度98%、特异度99%。
- dMRI与CBCT测量偏倚0.17 mm,95%一致性界限-1.05至1.38 mm。
- 不同FI分度的测量准确性无差异(p = .67)。
核心结论
- 与CBCT(无创金标准)相比,dMRI在评估上颌磨牙根分叉病变程度方面表现出高准确性和可靠性。
研究设计与样本
- 前瞻性研究,纳入22例重度牙周炎患者,对192个根分叉入口由两名独立观察者分别用CBCT和dMRI评估水平FI程度。
主要发现
- 基于CBCT,93个根分叉入口存在FI(I/II/III度分别为35/19/39个)。
- 两种模态的观察者内和观察者间一致性均极好(κ范围0.884至0.933)。
- dMRI测量与CBCT高度一致(偏倚0.17 mm;95%一致性界限-1.05至1.38 mm),且不同FI分度间测量准确性无差异(p = .67)。
- dMRI检测FI的敏感度和特异度分别为98%和99%。
- dMRI对FI分级的敏感度:I度89%、II度84%、III度100%。
临床怎么用
- dMRI可作为无辐射方式评估上颌磨牙根分叉病变程度,检测FI敏感度和特异度高。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIM: To investigate the accuracy and reliability of dental magnetic resonance imaging (dMRI) in assessing maxillary molar furcation involvement (FI). MATERIAL AND METHODS: In this prospective study, 22 patients with severe periodontitis underwent cone-beam computed tomography (CBCT) and dMRI. For 192 furcation entrances, the degree of horizontal FI was assessed by two independent observers on both modalities. Results of dMRI were compared with CBCT (reference modality) to assess the accuracy of dMRI. Cohen's kappa (κ), sensitivity and specificity were calculated for FI classification. Bland-Altman analysis and the Kruskal-Wallis test were used to evaluate measurement accuracy of dMRI. RESULTS: Based on CBCT findings, 93 furcation entrances revealed FI (degree I/II/III: 35/19/39). Intra- and inter-reader agreement was excellent for both modalities (κ-range: 0.884 to 0.933). dMRI measurements showed high agreement with CBCT (bias: 0.17 mm; 95% limits of agreement: -1.05 to 1.38 mm), and measurement accuracy did not differ among different degrees of FI (p = .67). For FI detection, sensitivity and specificity of dMRI were 98% and 99%. For FI classification, sensitivity values of dMRI were 89%/84%/100% for degree I/II/III. CONCLUSIONS: Compared to CBCT (non-invasive gold standard), dMRI demonstrates high accuracy and reliability in evaluating the degree of FI in maxillary molars.
引用
- Juerchott A, Sohani M, Schwindling FS, et al. In vivo accuracy of dental magnetic resonance imaging in assessing maxillary molar furcation involvement: A feasibility study in humans. J Clin Periodontol. 2020. PMID: 32343861. doi:10.1111/jcpe.13299
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。