四种镍钛锉在单根椭圆根管成形能力的微CT评价
结果显示所有锉均增加根管表面积、体积和锥度,组间无统计学差异(P<.05),且无一种锉能接触根管全部表面积。
要点
- 四种锉均增大椭圆根管表面积、体积和锥度,组间无显著差异。
- 无一种器械能完全接触椭圆根管全部表面积。
- 微CT可精确评估根管预备前后形态学变化。
核心结论
- WaveOne Gold、TRUShape、EdgeCoil和XP-3D Shaper预备椭圆根管后均相似地增加体积、表面积和锥度,且无一种系统能接触根管全部表面积。
主要发现
- 所有锉均增加根管表面积、体积和锥度(组间和组内)。
- 四种旋转器械组间任何指标均无统计学显著差异(P<.05)。
- 无一种锉系统能接触任何根管的全部表面积。
关键数值与推荐
- 样本量每组8颗牙;统计方法为单因素方差分析和配对样本t检验;显著性水平P<.05。
临床怎么用
- 四种锉对椭圆根管成形效果相似,临床选择可考虑其他因素如操作便利性,但均需注意椭圆根管壁的未接触区域。
局限性
- 体外研究,样本量较小(每组8颗),结果需临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study evaluated and compared the shaping ability of the WaveOne Gold (Dentsply/Tulsa Dental Specialties, Tulsa, OK), TRUShape 3D Conforming File (Dentsply/Tulsa Dental Specialties), EdgeCoil (EdgeEndo, Albuquerque, NM), and XP-3D Shaper (Brasseler USA, Savannah, GA) endodontic file systems on oval-shaped canals using micro-computed tomographic (micro-CT) technology. METHODS: Thirty-two oval-shaped, single-canal extracted human teeth were decoronated 1 mm coronal to the cementoenamel junction and scanned via a micro-CT scanner (μCT100; Scanco Medical, Bassersdorf, Switzerland). Teeth were divided into 4 groups (n = 8) and instrumented according to the manufacturer's instructions. Coregistered images, before and after root canal preparation, were evaluated for morphometric measurements of the surface area, volume, structure model index (SMI), conicity, and percent of walls untouched using the manufacturer's evaluation software (IPL Register, Scanco Medical). Data were statistically compared between groups using 1-way analysis of variance and within groups using a paired sample t test. RESULTS: Instrumentation with all file types increased the surface area, volume, and conicity between and within groups. There was no statistically significant difference between the groups for any of the rotary instruments used (P < .05). CONCLUSIONS: Instrumentation of oval-shaped canals with WaveOne Gold, TRUShape, EdgeCoil, and XP-3D Shaper rotary endodontic instruments similarly increase the volume, surface area, and conicity. None of the file systems were capable of contacting all of the surface area in any canal.
引用
- Thomas JP, Lynch M, Paurazas S, et al. Micro-computed Tomographic Evaluation of the Shaping Ability of WaveOne Gold, TRUShape, EdgeCoil, and XP-3D Shaper Endodontic Files in Single, Oval-shaped Canals: An In Vitro Study. J Endod. 2020. PMID: 32040423. doi:10.1016/j.joen.2019.11.001
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。