石墨烯纳米涂层抑制生物医用Ti-6Al-4V合金腐蚀
研究在Ti-6Al-4V 5级和23级圆片上制备石墨烯纳米涂层,并在0.5M NaCl加2-ppm氟化物、pH 2.0条件下腐蚀挑战至30天。结果显示涂层样品腐蚀速率低且稳定(约0.001 mm/y),未涂层对照末端腐蚀速率分别增至约0.091 mm/y的16倍和5倍,涂层阻止了表面氧化、降解和Ti、Al、V离子浸出。
要点
- 石墨烯涂层在30天腐蚀挑战中保持高于99%覆盖率。
- 涂层样品腐蚀速率约0.001 mm/y,未涂层对照约0.091 mm/y。
- 涂层阻止了未涂层对照中Ti、Al、V离子浸出。
核心结论
- 石墨烯涂层可有效保护Ti-6Al-4V免受腐蚀,为金属生物医用合金抗腐蚀涂层提供依据。
研究设计与样本
- 在Ti-6Al-4V 5级和23级圆片上制备石墨烯纳米涂层,并置于0.5M NaCl加2-ppm氟化物、pH 2.0条件下腐蚀挑战至30天。
主要发现
- 线性极化电阻曲线和电化学阻抗谱分析显示,石墨烯涂层样品在所有时间点均呈现更高耐腐蚀性和电化学稳定性。
- 石墨烯涂层样品腐蚀速率低且稳定(约0.001 mm/y),未涂层对照末端腐蚀速率分别增至约0.091 mm/y的16倍(5级)和5倍(23级)。
- 未涂层对照中观察到的表面氧化、降解(如缝隙缺陷)和Ti、Al、V离子浸出被石墨烯纳米涂层阻止。
- 拉曼映射确认石墨烯纳米涂层结构稳定性高、抗机械应力和化学降解,腐蚀挑战后覆盖率保持高于99%。
临床怎么用
- 石墨烯纳米涂层有望用于金属生物医用合金和植入器械的抗腐蚀保护。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The identification of metal ions and particles in the vicinity of failed implants has raised the concern that biomedical titanium alloys undergo corrosion in healthy and infected tissues. Various surface modifications and coatings have been investigated to prevent the deterioration and biocorrosion of titanium alloys but so far with limited success. Graphene is a cytocompatible atom-thick film made of carbon atoms. It has a very high surface area and can be deposited onto metal objects with complex shapes. As the carbon lattice has a very small pore size, graphene has promising impermeability capacity. Here, we show that graphene coating can effectively protect Ti-6Al-4V from corrosion. Graphene nanocoatings were produced on Ti-6Al-4V grade 5 and 23 discs and subjected to corrosive challenge (0.5M NaCl supplemented with 2-ppm fluoride, pH of 2.0) up to 30 d. The linear polarization resistance curves and electrochemical impedance spectroscopy analysis showed that the graphene-coated samples presented higher corrosion resistance and electrochemical stability at all time points. Moreover, the corrosion rate of the graphene-coated samples was very low and stable (~0.001 mm/y), whereas that of the uncoated controls increased up to 16 and 5 times for grade 5 and 23 (~0.091 mm/y) at the end point, respectively. The surface oxidation, degradation (e.g., crevice defects), and leaching of Ti, Al, and V ions observed in the uncoated controls were prevented by the graphene nanocoating. The Raman mappings confirmed that the graphene nanocoating presented high structural stability and resistance to mechanical stresses and chemical degradation, keeping >99% of coverage after corrosion challenge. Our findings open the avenues for the use of graphene as anticorrosion coatings for metal biomedical alloys and implantable devices.
引用
- Malhotra R, Han YM, Morin JLP, et al. Inhibiting Corrosion of Biomedical-Grade Ti-6Al-4V Alloys with Graphene Nanocoating. J Dent Res. 2020. PMID: 31905311. doi:10.1177/0022034519897003