根管入路角度与弯曲半径对镍钛旋转器械循环疲劳抗性的影响
本研究比较不同入路角度和弯曲半径对两种镍钛旋转器械循环疲劳抗性的影响,使用192支器械在16mm不锈钢人工根管中测试。结果显示小半径(3mm)和倾斜入路(10°、20°)会降低热 treated .04锥度器械的疲劳抗性,且HCM器械抗性高于TS,.04锥度高于.06锥度。
要点
- 小弯曲半径(3mm)比5mm显著降低器械循环疲劳抗性。
- 倾斜入路(10°、20°)显著降低.04锥度热 treated 器械的抗性。
- HCM器械循环疲劳抗性高于TS,.04锥度高于.06锥度。
核心结论
- 倾斜入路会降低所有测试半径下.04锥度热 treated 器械的循环疲劳抗性。
- 小弯曲半径与入路角度的协同效应会降低热 treated 器械的疲劳抗性。
研究设计与样本
- 体外研究,使用192支镍钛旋转器械(TS #25/.04、TS #25/.06、HCM #25/.04、HCM #25/.06)。
- 在16mm不锈钢人工根管(60°弯曲)中测试3种入路角度(0°、10°、20°)和2种半径(5mm、3mm)。
主要发现
- 3mm半径根管中器械循环疲劳抗性低于5mm半径(P < .05)。
- 3mm半径下,HCM #25/.06和所有.04锥度器械在20°和10°时NCF显著降低(P < .05),而TS2无显著差异。
- 5mm半径下,.06锥度器械各角度无显著NCF降低,.04锥度器械在20°时NCF显著降低(P < .05)。
- 相同尺寸器械比较,HCM的NCF高于TS(P < .05)。
- 相同热处理下,.04锥度器械NCF高于.06锥度(P < .05)。
临床怎么用
- 临床中应关注小弯曲半径和倾斜入路对镍钛器械疲劳抗性的不利影响。
- 对于热 treated .04锥度器械,倾斜入路可能增加疲劳断裂风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The aim of this study was to compare the influences from different access angles and curvature radii on cyclic fatigue resistance of nickel-titanium rotary files. METHODS: Two file systems (2Shape [TS; MicroMega, Besançon, France] and HyFlex CM [HCM; Coltène/Whaledent, Allstätten, Switzerland]) were used. A total of 192 instruments of TS #25/.04 (TS1), TS #25/.06 (TS2), HCM #25/.04, and HCM #25/.06 were evaluated at 3 insertion angles (0°, 10°, and 20°) and 2 radii (5 mm and 3 mm) in 16-mm stainless steel artificial canals with a 60° curvature. Cyclic fatigue resistance was determined by the number of cycles to failure (NCF) using a customized testing device. Data were analyzed statistically with the significance level established at 95%. RESULTS: In the 3-mm radius canal, the instruments showed lower cyclic fatigue resistance than in the 5-mm radius canal (P < .05). HCM #25/.06 and all .04 taper instruments had a significant NCF reduction at 20° and 10° in the 3-mm radius canal (P < .05), whereas TS2 showed no significant differences. In the 5-mm radius of curvature, although .06 taper instruments had no significant NCF reduction for each angle tested, .04 taper files exhibited significant NCF reduction when tested at 20° (P < .05). Comparing the same size instruments, HCM had higher NCF than TS (P < .05). Instruments with a .04 taper exhibited higher NCF than the .06 ones with the same heat treatment (P < .05). CONCLUSIONS: An inclined insertion into the canals decreased cyclic fatigue resistance of thermal-treated instruments with a .04 taper at all radii of curvature tested. The synergistic effect of a small radius of curvature and access angulation of heat-treated instruments decreases their fatigue resistance.
引用
- Pedullà E, La Rosa GRM, Virgillito C, et al. Cyclic Fatigue Resistance of Nickel-titanium Rotary Instruments according to the Angle of File Access and Radius of Root Canal. J Endod. 2020. PMID: 31911005. doi:10.1016/j.joen.2019.11.015