动态导航与徒手种植体植入精度的分口随机对照试验
该分口随机对照临床试验在32名志愿者上颌后牙区植入92枚种植体,比较基于微追踪器的动态导航与徒手方法植入种植体的偏差。最终分析86枚种植体,导航组在种植体肩部、根尖及角度偏差上均显著小于徒手组。结论认为导航技术能将虚拟种植计划更精确地转移到患者颌骨。
要点
- 动态导航辅助种植可提高植入精度,肩部线性偏差平均差0.72mm。
- 根尖线性偏差平均差0.69mm,角度偏差平均差5.33°。
- 分口随机对照设计,同一患者两侧上颌后牙区分别采用导航或徒手植入。
核心结论
- 导航技术可将虚拟种植计划以更高精度转移到患者颌骨。
研究设计与样本
- 分口随机对照临床试验,32名志愿者上颌后牙区每侧至少计划植入1枚种植体,随机采用动态导航或徒手植入。
- 共植入92枚种植体,86枚纳入最终分析,通过匹配计划数据与术后CBCT测量偏差,采用广义线性混合模型分析(p < .05)。
主要发现
- 种植体肩部线性偏差平均差0.72mm(SD 0.26;95%CI 0.39-1.02;p < .001)。
- 种植体根尖线性偏差平均差0.69mm(SD 0.36;95%CI 0.19-1.19;p < .001)。
- 角度偏差平均差5.33°(SD 1.63;95%CI 7.17-3.48;p < .001)。
临床怎么用
- 对于上颌后牙区种植,可考虑使用基于微追踪器的动态导航设备以提高植入精度。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: The aim of this split-mouth randomized controlled clinical trial was to compare the deviations of planned and placed implants placed by the assistance of a micron tracker-based dynamic navigation device or freehand methods. MATERIAL AND METHODS: A thermoplastic fiducial marker was adapted on the anterior teeth, and cone-beam computerized tomography was used for imaging. A minimum of one implant was planned for each side of the posterior maxilla, and the dynamic navigation device or freehand method was randomly used for surgical insertion. Deviations were measured by matching the planning data with a final CBCT image. Linear deviations (mm) between the planned and placed implants were the primary outcome. The results were analysed by generalized linear mixed models (p < .05). (NCT03471208). RESULTS: A total of 92 implants were placed to 32 volunteers, and 86 implants were included in the final analysis. For the linear deviations, mean of differences (Δ) was 0.72mm (Standard deviation (SD): 0.26); (95% Confidence interval (CI): 0.39-1.02) in the shoulder of the implants (p < .001) and 0.69mm (SD: 0.36); (95% CI: 0.19-1.19) in the tip of the implants (p < .001). For the angular deviations, Δ was 5.33° (SD: 1.63); (95% CI: 7.17-3.48); (p < .001). CONCLUSIONS: The navigation technique can be used to transfer virtual implant planning to the patient's jaw with increased accuracy.
引用
- Aydemir CA, Arısan V. Accuracy of dental implant placement via dynamic navigation or the freehand method: A split-mouth randomized controlled clinical trial. Clin Oral Implants Res. 2019. PMID: 31829457. doi:10.1111/clr.13563
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。