多次使用Reciproc Blue器械在再治疗中的循环疲劳抗力
本研究评估Reciproc Blue R25器械在下颌磨牙再治疗中多次使用对循环疲劳抗力的影响。结果显示新器械与首次使用器械的NCF无显著差异,但均显著高于第二次和第三次使用器械;第二次与第三次使用间无差异;再治疗时间随使用次数增加而延长。结论认为Reciproc Blue R25可用于去除2颗磨牙的根管充填材料,但超过2颗磨牙的多次使用可能不建议。
要点
- 新器械与首次使用器械的循环疲劳抗力无显著差异。
- 第二次和第三次使用器械的循环疲劳抗力显著低于新器械和首次使用器械。
- 再治疗操作时间随器械使用次数增加而显著延长。
核心结论
- Reciproc Blue R25器械可用于去除2颗磨牙的根管充填材料,但超过2颗磨牙的多次使用可能因NCF值下降而不建议。
主要发现
- 新器械与首次使用器械的NCF值无显著差异(P > .05)。
- 新器械和首次使用器械的NCF值显著高于第二次和第三次使用器械(P < .05)。
- 第二次和第三次使用器械的NCF值无显著差异(P > .05)。
- 再治疗操作的平均时间随使用次数增加而显著延长(P < .05)。
临床怎么用
- Reciproc Blue R25可用于去除2颗磨牙的根管充填材料。
- 超过2颗磨牙的多次使用可能因循环疲劳抗力下降而不建议。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Root canal retreatment procedure may reduce the mechanical resistance of nickel-titanium instruments and increase the frequency of instrument fracture. The aim of the present study was to evaluate the effect of multiuse of Reciproc Blue R25 instruments on cyclic fatigue resistance during retreatment procedure in mandibular molar teeth. METHODS: Forty-eight Reciproc Blue R25 files were included in the study. In control group, 12 new Reciproc Blue R25 instruments were subjected to cyclic fatigue test by using a stainless steel artificial canal with a 90° angle and 3-mm radius of curvatures. In experimental groups, Reciproc Blue R25 files (n = 12) were used to remove the root canal filling (RCF) material of 1, 2, and 3 molar teeth, respectively. Then cyclic fatigue tests of these 36 instruments were performed. The number of cycles to fracture (NCF) and the length of the fractured fragment were calculated for each instrument. Working time for retreatment procedures was determined in seconds with a digital chronometer. The data were analyzed by using one-way analysis of variance followed by post hoc Tukey test and mixed-effect analysis (P = .05). RESULTS: No significant difference was found between NCF values of new and first time used files (P > .05), but NCF values of new and first time used files were significantly higher than those of second and third time used files (P < .05). The second and third time used files revealed no difference (P > .05). As the number of uses increased, the average time for retreatment procedure statistically extended (P < .05). CONCLUSIONS: Reciproc Blue R25 instrument can be used to remove RCF material of 2 molar teeth. However, because of its decreased NCF values, multiple use over 2 molars may not be suggested.
引用
- Serefoglu B, Miçooğulları Kurt S, Kaval ME, et al. Cyclic Fatigue Resistance of Multiused Reciproc Blue Instruments during Retreatment Procedure. J Endod. 2019. PMID: 31812359. doi:10.1016/j.joen.2019.10.024
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。