PEEK与钛种植体封闭帽的软组织反应比较

Soft tissue response to dental implant closure caps made of either polyetheretherketone (PEEK) or titanium.
PubMed 文献 · Clin Oral Implants Res · 2019-06-12

组织形态学显示两种材料均无炎症征象,但PEEK封闭帽表面MNGCs数量显著高于钛封闭帽。

要点

  • PEEK封闭帽表面的多核巨细胞数量显著多于钛封闭帽。
  • 两种封闭帽及三种种植体均未见组织学炎症征象。
  • 三种种植体类型间MNGCs数量无统计学显著差异。

核心结论

  • PEEK封闭帽接触的多核巨细胞显著多于钛封闭帽。

研究设计与样本

  • 采用组织形态学方法评估不同区域接触PEEK或钛的小型与大型多核巨细胞数量。

主要发现

  • PEEK与钛封闭帽及三种种植体上均观察到多核巨细胞,未见组织学炎症征象。
  • PEEK封闭帽在面向软组织和硬组织的外表面及内表面MNGCs数量均显著高于钛封闭帽。
  • 三种种植体类型间MNGCs数量无统计学显著差异。

关键数值与推荐

  • PEEK位于氧化锆上时外表面面向软组织p=0.0008、面向硬组织p=0.016、内表面p=0.014。
  • PEEK位于氧化铝增韧氧化锆上时外表面面向软组织p=0.0016、面向硬组织p=0.003、内表面p=0.0088。

临床怎么用

  • 选择种植体封闭帽材料时可考虑PEEK与钛在软组织反应上的差异,关注多核巨细胞反应。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Polyetheretherketone (PEEK) is a popular synthetic thermoplastic polymer for medical applications, but its clinical use suffers from several limitations. Therefore, the aim was to compare the soft tissue response to dental implant closure caps made of PEEK or titanium as evaluated by the occurrence of multinucleated giant cells (MNGCs). MATERIAL AND METHODS: Forty-two implants were placed in the maxilla of seven miniature pigs. While commercially pure titanium (Ti) implants had a Ti closure cap, ceramic implants made of either zirconia (Zr) or alumina-toughened zirconia (Zr + Al) received a PEEK closure cap. Histomorphometry was performed to evaluate the number of small and large MNGCs being in contact with the PEEK or the Ti in different compartments of the implant systems. RESULTS: No histological signs of inflammation were noticed, and MNGCs were observed on both PEEK and Ti closure caps and on all three implant types. Significantly higher numbers of MNGCs were found on closure caps made of PEEK than on closure caps made of Ti on the external closure cap surface facing both soft (p = 0.0008 for PEEK on Zr and p = 0.0016 for PEEK on Zr + Al) and hard tissues (p = 0.016 for PEEK on Zr and p = 0.003 for PEEK on Zr + Al) as well as in the internal closure cap surface (p = 0.014 for PEEK on Zr and p = 0.0088 for PEEK on Zr + Al). No statistically significant differences in the number of MNGCs were observed on the three implant types. CONCLUSIONS: Significantly more MNGCs were in contact with PEEK than with Ti closure caps.

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。

引用

  1. Caballé-Serrano J, Chappuis V, Monje A, et al. Soft tissue response to dental implant closure caps made of either polyetheretherketone (PEEK) or titanium. Clin Oral Implants Res. 2019. PMID: 31125452. doi:10.1111/clr.13487

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。