库存与CAD/CAM定制氧化锆种植基台临床功能1年后的相变与断裂载荷

Phase transformation and fracture load of stock and CAD/CAM-customized zirconia abutments after 1 year of clinical function.
PubMed 文献 · Clin Oral Implants Res · 2019-05-03

本研究比较了23个库存和23个CAD/CAM定制氧化锆种植基台临床使用1年后与原始对照的断裂强度和单斜相氧化锆体积百分比。结果显示库存基台断裂载荷显著降低至78.8%,CAD/CAM基台无降低,所有样本均未检测到超过5%的单斜相。

要点

  • 库存氧化锆基台临床使用1年后断裂载荷显著降低
  • CAD/CAM定制氧化锆基台断裂载荷未见降低
  • 未观察到明显的四方相到单斜相转变

核心结论

  • 临床使用1年后,CAD/CAM定制氧化锆基台的断裂强度无差异,而库存氧化锆基台断裂强度显著降低。
  • 未观察到实质性的四方相到单斜相转变。

研究设计与样本

  • 离体实验,比较23个库存和23个CAD/CAM定制氧化锆种植基台临床使用1年后与几何相同的原始对照。
  • 使用拉曼光谱测定单斜相体积分数,单载荷至断裂试验测定断裂载荷,光学和SEM显微镜进行失效分析。
  • 正态性验证后使用配对t检验比较断裂载荷,显著性水平α=0.05。

主要发现

  • 库存氧化锆基台临床功能1年后断裂载荷显著降低至78.8%(SD 29.5%),p<0.05。
  • CAD/CAM基台未发现断裂载荷降低。
  • 所有样本中未观察到超过检测阈值5%的单斜相氧化锆体积百分比。

临床怎么用

  • 库存氧化锆基台临床使用后断裂强度可能下降,需关注长期性能。
  • CAD/CAM定制氧化锆基台在1年临床功能后断裂强度保持稳定。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Functional loading and low-temperature degradation may give rise to impaired clinical long-term service of zirconia implant abutments. The aim of this study was to compare the fracture strength (primary outcome measure) and the volume percentage of monoclinic surface zirconia (m-ZrO2 ) of stock and CAD/CAM-customized zirconia implant abutments that functioned clinically for 1 year with geometrically identical pristine controls in an ex vivo experiment. MATERIAL AND METHODS: Twenty-three stock (ZirDesign™) and 23 CAD/CAM-customized (Atlantis™) zirconia implant abutments were retrieved after 1 year of clinical service. They were compared with pristine copies with respect to the volume fraction of the monoclinic phase using Raman spectroscopy and their fracture load by means of a single load-to-fracture test. Failure analysis was performed using optical and SEM microscopy. After verification of normal distribution, paired t tests were used for comparison of fracture loads between pristine and clinically aged specimen. All statistical tests employed a level of significance of α = 0.05. RESULTS: The fracture loads of the stock zirconia abutments were significantly (p < 0.05) reduced to 78.8% (SD 29.5%) after one year of clinical function. For the CAD/CAM abutments, no reduction in fracture load was found. No m-ZrO2 volume percentages beyond the detection threshold of 5% were observed in any of the samples. CONCLUSIONS: After 1 year of clinical service, no difference in fracture strength of the CAD/CAM-customized zirconia implant abutments could be demonstrated, whereas the stock zirconia abutments decreased considerably in fracture strength. No substantial tetragonal-to-monoclinic transformation was observed.

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。

引用

  1. Schepke U, Gresnigt MMM, Browne WR, et al. Phase transformation and fracture load of stock and CAD/CAM-customized zirconia abutments after 1 year of clinical function. Clin Oral Implants Res. 2019. PMID: 31009128. doi:10.1111/clr.13442

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。