种植体断裂失败率及潜在相关风险指标:5124例患者最长12年回顾性研究
Kaplan-Meier分析显示3年、5年、10年种植体生存率分别为99.8%、99.2%、97.7%。多因素Cox回归显示种植体直径、位置、植骨史和微螺纹与断裂显著相关。
要点
- 内连接种植体断裂发生率为0.92%,10年生存率97.7%。
- 宽直径种植体在90个月内断裂风险降低,之后直径与断裂无关。
- 无植骨史或无微螺纹的种植体在整个随访期更易断裂。
核心结论
- 内连接种植体断裂发生率较低,直径、位置、植骨史和微螺纹是断裂的显著相关风险指标。
研究设计与样本
- 回顾性分析2006至2015年间在退伍军人健康服务医疗中心牙科医院植入的19006枚内连接种植体,涉及5124例患者(4570例男性,554例女性),随访至2018年6月。
主要发现
- Kaplan-Meier估计的3年、5年、10年种植体生存率分别为99.8%、99.2%、97.7%。
- 多因素Cox回归显示种植体直径、位置、植骨史和微螺纹与断裂显著相关。
- 宽直径种植体在90个月内断裂风险降低,90个月后直径与断裂不相关。
- 前下颌种植体在90个月内断裂风险较低,下颌前磨牙区种植体在90个月后风险较低。
临床怎么用
- 临床选择内连接种植体时,应关注直径、植入位置、植骨史和微螺纹设计以降低断裂风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: This study investigated fracture rates and risk indicators for fractures in internal connection dental implants. MATERIAL AND METHODS: We performed a retrospective analysis of 19,006 internal connection implants used in fixed restoration in 5,124 patients (4,570 males, 554 females) at the Dental Hospital of Veterans Health Service Medical Center between 2006 and 2015. Patients were followed through June 2018 (0.03-12.39 years post-installation). Clinical factors (age, sex, implant diameter, implant length, placement site, bone graft, fixture material, cervical feature, abutment connection, microthread, and platform switching) were recorded. Kaplan-Meier survival analysis identified risk indicators associated with an implant fracture. Cox regression models elucidated potential fracture risks. RESULTS: One hundred and seventy-four implants fractured in 135 patients, for an incidence rate of 0.92% after an average of 4.95 ± 2.14 years of use. Kaplan-Meier estimates showed that the 3-, 5-, and 10-year survival rates of implants were 99.8%, 99.2%, and 97.7%, respectively. In the multivariable Cox regression model, the diameter, location, history of bone graft, and microthread presence were significantly correlated with implant fractures. Wide-diameter implants had a reduced fracture risk within 90 months, after which the diameter did not correlate with fractures. Implants placed in the anterior mandible had a lower fracture risk within 90 months; mandibular premolar implants corresponded with a lower risk after 90 months. Implants without a history of bone graft or microthreads were more likely to fracture throughout the follow-up time. CONCLUSIONS: These results elucidate risk indicators for implant fractures and facilitate their reduction in clinical practice.
引用
- Lee DW, Kim NH, Lee Y, et al. Implant fracture failure rate and potential associated risk indicators: An up to 12-year retrospective study of implants in 5,124 patients. Clin Oral Implants Res. 2019. PMID: 30672029. doi:10.1111/clr.13407
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。