原厂与非原厂基台用于螺丝固位单颗种植冠的体外评估
原厂组冠-基台间隙显著小于非原厂组,原厂基台-种植体螺丝扭矩衰减百分比更低,但各组承载能力差异无统计学意义。
要点
- 原厂组件冠-基台间隙显著小于非原厂组件。
- 原厂基台-种植体螺丝循环加载后扭矩衰减百分比更低。
- 各组静态承载能力差异无统计学意义。
核心结论
- 使用原厂组件可获得更好的适合性。
- 原厂基台在循环加载后扭矩衰减百分比低于非原厂基台。
研究设计与样本
- 内部适合性用扫描电镜评估每组8个样本的截面。
- 承载能力按ISO 14801标准对每组8个样本加载至断裂。
- 循环加载(2×10^6次)前后评估每组5个样本的螺丝移除扭矩值。
主要发现
- 冠-基台平均间隙:原厂组9.3 μm,非原厂组分别为45.4 μm和44.9 μm,组间差异有统计学意义。
- 平均承载能力:原厂组1098 N,非原厂组分别为1057 N和973 N,差异无统计学意义。
- 原厂基台-种植体螺丝的扭矩衰减百分比低于非原厂。
关键数值与推荐
- 冠-基台间隙原厂组9.3 μm,非原厂组45.4 μm和44.9 μm,差异有统计学意义(Kruskal-Wallis检验,p < 0.05)。
- 平均承载能力原厂组1098 N,非原厂组1057 N和973 N,差异无统计学意义。
临床怎么用
- 对于螺丝固位单颗种植冠,选用原厂组件可能改善内部适合性并减少螺丝扭矩衰减。
局限性
- 本研究为体外研究,结果需临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To assess the internal accuracy, mechanical behaviour under static load and screw loosening before and after cyclic loading of implant-supported crowns restored with original components or with two compatible non-original brands. MATERIALS AND METHODS: Sixty-three dental implants were divided into three groups (n = 21 each): Group 1 used original components, and groups 2 and 3 used non-original components. Internal accuracy was measured throughout a cross section of the sample groups (n = 8) using scanning electron microscopy (SEM) to evaluate the fit for implant-abutments, implant-crowns and crown-abutments. To evaluate load-bearing capacity, eight samples from each group were loaded until fracture according to ISO Norm 14801. The removal torque value (RTV) was evaluated in the screws connecting the crown with the abutment and the abutment with the implant before and after cyclic loading (2 x 106 cycles) in five samples from each group. The data were analysed using the Kruskal-Wallis test (p < 0.05). RESULTS: Statistically significant differences were found among the mean crown-abutment gaps measured in Group 1 (9.3 μm), Group 2 (45.4 μm) and Group 3 (44.9 μm). Higher values for mean load-bearing capacity were found in Group 1 (1,098 N) than in groups 2 and 3 (1,057 and 973 N, respectively); however, these differences were not statistically significant. Original abutment-implant screws exhibited lower percentages of torque reduction than the non-originals. CONCLUSIONS: Enhanced fit is expected when original components are used. The original abutments exhibited lower percentages of torque reduction after cyclic loading than non-originals.
引用
- Alonso-Pérez R, Bartolomé JF, Ferreiroa A, et al. Original vs. non-original abutments for screw-retained single implant crowns: An in vitro evaluation of internal fit, mechanical behaviour and screw loosening. Clin Oral Implants Res. 2018. PMID: 30537302. doi:10.1111/clr.13390
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。