高分辨率小视野CBCT检测根尖峡部复杂解剖的准确性:离体分析
本研究以显微CT为参考标准,评估小视野CBCT检测下颌磨牙根尖峡部的准确性。结论认为小视野CBCT不能准确检测和测量根尖峡部长度。
要点
- 小视野CBCT检测下颌磨牙根尖峡部敏感度为65%(95%CI 0.4667-0.8333)。
- 仅74.7%的峡部长度可被CBCT测量,且与显微CT长度差异显著(P < .05)。
- 高分辨率CBCT不能可靠预测实际根尖根管解剖。
核心结论
- 小视野CBCT成像不能准确检测和测量下颌磨牙根尖峡部长度,高分辨率设置亦不能可靠预测实际根尖根管解剖。
主要发现
- CBCT检测峡部的敏感度为65%(95%CI 0.4667-0.8333)。
- 平均74.7%的峡部长度可被测量。
- 显微CT与CBCT测量的峡部长度差异显著(P < .05;均值0.756 ± 0.655;t检验),两种方法无一致性。
关键数值与推荐
- 线性回归用于显微CT真实长度与CBCT长度之间分析,Bland-Altman图及t检验,α = 0.05。
临床怎么用
- 临床不应依赖小视野CBCT高分辨率图像来准确判断或测量下颌磨牙根尖峡部的实际解剖。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The purpose of this study was to evaluate the accuracy of small-volume cone-beam computed tomography (CBCT) to detect and measure isthmi in the apical root canals of mandibular molars by using micro-computed tomography (μ-CT) as the reference standard. METHODS: Forty mandibular first molars selected on the basis of μ-CT scan and presenting isthmi in the apical 3-mm mesial roots were scanned by using the highest-resolution settings of a small-volume CBCT unit. Isthmi lengths were measured and compared between both μ-CT and CBCT images to study the accuracy of CBCT readings. Quantitative data for sensitivity rate were depicted as percentage value with 95% confidence interval. Results were analyzed by using linear regression between true lengths (μ-CT) and CBCT lengths, Bland-Altman plot and t test, at α = 0.05. RESULTS: CBCT sensitivity for isthmi detection was 65% (95% confidence interval, 0.4667-0.8333). An average of 74.7% of the lengths could be measured, and differences among the lengths in μ-CT and CBCT were significant (P < .05; mean, 0.756 ± 0.655; t test), showing that there was no agreement between both methods. CONCLUSIONS: Accuracy of identifying apical isthmi of mandibular molars was highly influenced by the evaluation method. Small-volume CBCT imaging could not detect and measure apical isthmi length accurately. Moreover, using high-resolution settings in CBCT, it was not reliable to forecast the actual apical root canal anatomy.
引用
- Tolentino ES, Amoroso-Silva PA, Alcalde MP, et al. Accuracy of High-resolution Small-volume Cone-beam Computed Tomography in Detecting Complex Anatomy of the Apical Isthmi: Ex Vivo Analysis. J Endod. 2018. PMID: 30390974. doi:10.1016/j.joen.2018.08.015
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。