文献速递牙体牙髓病

微型CT评估3种根尖定位仪在下颌磨牙弯曲根管中的准确性

A Micro-computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars.
PubMed 文献 · J Endod · 2018-11-01

本研究用微型CT扫描评估CanalPro、Apex ID和Root ZX Mini三种根尖定位仪在离体下颌磨牙弯曲近中根管中的准确性。结果显示Root ZX Mini和CanalPro在根管长度和工作长度测定中均精确,而Apex ID仅在0.5标记时工作长度准确。

要点

  • Root ZX Mini和CanalPro在弯曲根管中根管长度及工作长度测定均精确。
  • Apex ID的电子根管长度与金标准一致性差(P < .05),但0.5标记工作长度准确。
  • 侧方根尖孔对Apex ID的0.0标记准确性有负面影响(P < .05)。

核心结论

  • Root ZX Mini和CanalPro在离体下颌磨牙弯曲近中根管中根管长度和工作长度测定均精确。
  • Apex ID仅在0.5标记时工作长度测定准确,其电子根管长度准确性较差。

研究设计与样本

  • 离体下颌磨牙58个弯曲近中根管,用视觉法和微型CT三维重建测定实际根管长度和工作长度。
  • 记录三种根尖定位仪在APEX/0.0和0.5标记的电子测量值,与金标准比较。

主要发现

  • 除Apex ID的电子根管长度外(P < .05),所有电子测量与各自金标准高度一致。
  • 三种设备在±0.5 mm精确测量百分比上无差异(χ2和z检验,P > .05)。
  • 侧方根尖孔的存在对Apex ID的0.0标记有负面影响(Wilcoxon检验,P < .05)。

关键数值与推荐

  • 精确测量定义为电子测量值与实际长度偏差在±0.5 mm以内。
  • 电子根管长度取APEX/0.0标记读数,电子工作长度取0.5标记读数。

临床怎么用

  • 在下颌磨牙弯曲近中根管中可优先选用Root ZX Mini或CanalPro进行根管长度和工作长度测定。
  • 使用Apex ID时建议以0.5标记确定工作长度,并注意侧方根尖孔可能影响0.0标记准确性。

局限性

  • 研究为离体牙实验,结果需临床验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: This study evaluated the accuracy of 3 electronic apex locators (EALs) (CanalPro [Coltene-Endo, Cuyahoga Falls, OH], Apex ID [SybronEndo, Glendora, CA], and Root ZX Mini [J Morita, Tokyo, Japan]) in curved mesial canals of extracted mandibular molars using micro-computed tomographic (micro-CT) scanning. METHODS: The root canal length and the actual working length of 58 canals were measured using the visual method and 3-dimensional micro-CT reconstructions. The measurements of the EALs at marks "APEX/0.0" and "0.5 mark" were recorded as the electronic root canal length and the electronic working length, respectively. The absolute mean values and the percentages of distribution of the electronic measurements were compared with the actual lengths of the canals. The presence of isthmuses, accessory canals, foramen position, and the distance between the apical constriction and the apical foramen were recorded from the micro-CT scans, and their correlation to the accuracy of the EALs was determined. RESULTS: All electronic measurements showed high agreement with their respective gold standard, except the electronic root canal length of the Apex ID (P < .05). No difference in the percentage of precise measurements (within the ±0.5 mm) was found among the devices (χ2 and z tests, P > .05). Of the anatomic parameters evaluated, the presence of a lateral foramen negatively affected the 0.0 mark of the Apex ID (Wilcoxon test, P < .05). CONCLUSIONS: The Root ZX Mini and CanalPro were precise for both root canal length and working length determination in mesial curved canals of mandibular molars, whereas the Apex ID was accurate for the working length when using the 0.5 mark.

来源:J Endod,查看原文。版权归原作者所有。

引用

  1. Piasecki L, José Dos Reis P, Jussiani EI, et al. A Micro-computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars. J Endod. 2018. PMID: 30390968. doi:10.1016/j.joen.2018.09.001

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。