体内根管器械操作扭矩测量:啄法与刷法对比

Measurement of torque generated during intracanal instrumentation in vivo.
PubMed 文献 · Int Endod J · 2018-12-04

比较向内向下的啄法与向外侧向的刷法,结果显示刷法所需扭矩显著更低(P < 0.05),但两种操作均未超过设定扭矩上限。

要点

  • 体内测量显示刷法(outward action)比啄法(inward action)所需扭矩显著更低。
  • 两种操作模式下记录扭矩值均未超过2.5 N cm的设定上限。
  • 操作技术显著影响器械所受扭矩,提示临床操作方式可影响器械负荷。

核心结论

  • 体内根管器械操作时,向外刷法比向内啄法所需扭矩显著更低,操作技术显著影响器械所受扭矩。
  • 两种操作均未达到设定的2.5 N cm扭矩上限。

研究设计与样本

  • 比较向内啄法与向外刷法,采用t检验和Wilcoxon检验,显著性水平设为5%。

主要发现

  • 向内啄法与向外刷法在评估的扭矩参数上差异显著(P < 0.05)。
  • 向外刷法所需扭矩显著更低,被认为比向内啄法更安全。
  • 所有记录扭矩值均未超过设定的2.5 N cm扭矩上限。

临床怎么用

  • 临床使用旋转器械时可考虑采用向外刷法以降低扭矩负荷。
  • 体内扭矩监测有助于评估不同操作技术的器械负荷差异。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: To compare two different clinical techniques when using endodontic rotary instruments by monitoring the torque in vivo. METHODOLOGY: Ten single-rooted maxillary anterior teeth were prepared by an endodontist using size 25, .06 taper TF rotary instruments (KerrEndo, Orange, CA, USA). All instruments were rotated at 500 rpm with maximum torque set at 2.5 N cm using an endodontic motor (Kavo, Biberach, Germany), which automatically recorded the torque values every 1/10 of a second. Instruments were used with two modi operandi: (i) Inward action: short-length intermittent progression with slight apical pressure aiming to reach the working length (also known as pecking motion), followed by (ii) Outward action: lateral brushing action selectively directed to address the canal circumference whilst removing instruments in the coronal direction. The mean, maximum and accumulated torque values were recorded and analysed statistically (t-test and Wilcoxon test) with a significance level set at 5%. None of the recorded values exceeded the selected torque limit. RESULTS: A significant difference between inward and outward actions was found regarding the evaluated parameters (P < 0.05). The outward action (brushing) was significantly safer (requiring less torque) compared to the inward action (pecking). CONCLUSIONS: Torque measurements during in vivo instrumentation provided useful information regarding the techniques evaluated. Although the torque limit was not reached by any of the two actions, the technique impacted significantly on the torque applied to the instrument.

来源:Int Endod J,查看原文。版权归原作者所有。