钛表面生物活性玻璃功能化对骨结合的影响:犬实验研究
结果显示BF处理未改变表面粗糙度但降低接触角,2周时BF组骨-种植体接触和骨密度显著高于对照组,ISQ值两组无显著差异。
要点
- BSF18功能化降低种植体表面接触角,改善润湿性。
- 2周时BF组BIC和BD显著高于对照组。
- 两组ISQ值无显著差异,4周时均高于基线。
核心结论
- BSF18功能化改善了种植体表面润湿性,并在2周时提高了骨-种植体接触和骨密度。
研究设计与样本
- 种植体在愈合2周和4周后进行分析,检测ISQ、BIC和BD,BF组另做原子力显微镜粗糙度分析和接触角分析。
- 数据采用双因素方差分析和Tukey检验(p < 0.05)。
主要发现
- BF处理未影响表面粗糙度,BF组接触角低于对照组。
- 两组种植体4周时ISQ值均显著高于基线(p = 0.04)。
- 2周时BF组BIC(p = 0.011)和BD(p = 0.025)显著高于对照组。
- 对照组4周时BIC(p = 0.030)和BD(p = 0.015)显著高于2周时,BF组2周与4周间无显著差异。
临床怎么用
- BSF18功能化可能通过改善润湿性促进早期骨结合,但需进一步临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The purpose of this study was to evaluate the effect of surface functionalization with bioactive glass BSF18 on the osseointegration of sandblasted and dual acid-etched surface (AE) implants. METHODS AND MATERIALS: Forty Morse taper implants with an AE surface as controls (C) or with an AE surface functionalized with BSF18 (BF) were placed in the mandibles of 10 beagles. Implants were analyzed after 2 and 4 weeks of healing. Implant stability quotient (ISQ) values were registered immediately after installation and prior to sacrifice. Samples were analyzed for bone-to-implant contact (BIC) and bone density (BD). The characterization of BF implants included surface roughness analysis with atomic force microscopy and contact angle (CA) analysis to evaluate wettability. Data were analyzed using two-way ANOVA followed by Tukey's test (p < 0.05). RESULTS: Surface roughness was not affected by BF treatment. CA was lower in the BF group compared to the C group. No significant difference was observed in ISQ values between surfaces (p = 0,231), irrespective of time. Significantly higher ISQ values were observed for both implants after 4 weeks when compared with baseline (p = 0.04). Significantly higher BIC (p = 0.011) and BD (p = 0.025) values were observed for the BF compared to the C group at 2 weeks. Significantly higher BIC (p = 0.030) and BD (p = 0.015) values for the C group were observed at 4 weeks compared to 2 weeks. No significant difference was observed in the BF group between 2 and 4 weeks. CONCLUSIONS: Implant functionalization with BSF18 improved the wettability of the implant surface; enhancing BIC and BD at 2 weeks.
引用
- Soares PBF, Moura CCG, Chinaglia CR, et al. Effect of titanium surface functionalization with bioactive glass on osseointegration: An experimental study in dogs. Clin Oral Implants Res. 2018. PMID: 30264907. doi:10.1111/clr.13375
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。