Twisted File Adaptive与Navigator EVO在不同运动模式下的根管偏移、居中能力和循环疲劳
本研究比较Twisted File Adaptive和Navigator EVO系统在自适应与连续旋转运动下的循环疲劳和根管偏移。两种系统均可用于两种运动,但连续旋转下器械寿命更高,根管偏移较小。
要点
- 两种系统在自适应和连续旋转下均可使用,根尖3和9 mm处偏移和居中能力相似。
- 连续旋转下器械循环疲劳抗力高于自适应运动,与系统和器械尖端/直径无关。
- 上颌磨牙近中颊根预备至35.04时仅发生小量根管偏移。
核心结论
- 两种系统均可用于自适应或连续旋转,但连续旋转下器械寿命更高,上颌磨牙近中颊根预备至35.04时仅发生小量根管偏移。
研究设计与样本
- 采用Kruskal-Wallis和Dunn检验评估根管偏移、居中能力和根管体积,Student t检验评估循环疲劳(P = .05)。
主要发现
- 距根尖3和9 mm处,所有组根管偏移和居中能力相似(P > .05)。
- 距根尖6 mm处,TFC向根分叉区偏移高于NA和NC(P < .05)。
- 预备后TFA牙本质去除更多,根管体积高于NA和NC(P < .05)。
- 无论系统或器械尖端/直径,连续旋转下循环疲劳抗力均高于自适应运动(P < .05)。
临床怎么用
- 临床可选用TF Adaptive或Navigator EVO系统,运动模式可根据器械寿命需求选择连续旋转。
- 上颌磨牙近中颊根预备至35.04时根管偏移较小,但仍需关注距根尖6 mm区域偏移。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study compared the cyclic fatigue and the canal transportation promoted by Twisted File (TF) Adaptive and Navigator EVO systems when used with 2 different motions. METHODS: Forty mesiobuccal roots of maxillary molars were scanned by using micro-computed tomography imaging before and after root canal preparation with the 2 instrument systems used with 2 motions (adaptive and continuous rotation). Samples were divided into 4 groups: TFA, TF Adaptive instruments under adaptive motion; TFC, TF Adaptive instruments under continuous motion; NA, Navigator instruments under adaptive motion; and NC, Navigator instruments under continuous motion. Root canals were prepared until 35.04 instruments. Apical transportation was analyzed by using micro-computed tomography at 3 levels: 3, 6, and 9 mm from the apex. The cyclic fatigue tests were performed by using a custom-made device. Ten instruments of each brand were activated by using a 6:1 reduction handpiece powered by a torque-controlled motor using the preset programs "custom mode" and "TF Adaptive" to activate 25.06 and 35.04 instruments. Kruskal-Wallis and Dunn tests were used to assess canal transportation, centering ability, and canal volume. The Student t test was used to evaluate cyclic fatigue (P = .05). RESULTS: At 3 and 9 mm, the canal transportation and centering ability were similar in all groups (P > .05). At 6 mm, TFC presented higher canal transportation toward furcal region than NA and NC (P < .05). After canal preparation, TFA promoted great dentinal excision, presenting higher canal volume than NA and NC (P < .05). Higher cyclic fatigue resistance was observed under continuous than adaptive motion regardless of system or tip/diameter of the instrument (P < .05). CONCLUSIONS: Both systems can be used under adaptive or continuous rotation. However, the life span of the instruments was higher when used under continuous rotation. Small canal transportation occurred when mesiobuccal root canals from maxillary molars were prepared until 35.04 instruments.
引用
- Marks Duarte P, Barcellos da Silva P, Alcalde MP, et al. Canal Transportation, Centering Ability, and Cyclic Fatigue Promoted by Twisted File Adaptive and Navigator EVO Instruments at Different Motions. J Endod. 2018. PMID: 30078572. doi:10.1016/j.joen.2018.06.002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。