预备滑行通道对Reciproc与Reciproc Blue镍钛锉循环疲劳抗力的影响:一项实验室研究
本研究比较了Reciproc与Reciproc Blue镍钛锉在有无滑行通道条件下预备下颌磨牙根管后的循环疲劳抗力。共使用60根Reciproc R25和60根Reciproc Blue R25锉,分为未使用、无滑行通道使用和有滑行通道使用三组,在模拟根管中旋转至折断。结果显示Reciproc Blue的循环疲劳抗力高于Reciproc,但使用ProGlider创建滑行通道对两种锉的循环疲劳抗力均无显著影响。
要点
- Reciproc Blue镍钛锉的循环疲劳抗力高于Reciproc镍钛锉。
- 使用ProGlider创建滑行通道未改变Reciproc和Reciproc Blue的循环疲劳抗力。
- 两种锉折断片段平均长度无统计学显著差异。
核心结论
- 在本体外研究限制内,使用ProGlider锉创建滑行通道对Reciproc和Reciproc Blue镍钛锉的循环疲劳抗力无影响。
研究设计与样本
- 实验室研究,使用60根Reciproc R25和60根Reciproc Blue R25镍钛锉。
- 每种锉分为3个亚组:原厂状态、无滑行通道使用、有滑行通道使用。
- 所有器械在内径1.5 mm、弯曲角度60°、弯曲半径5 mm的不锈钢人工根管中旋转至折断。
主要发现
- 原厂状态Reciproc Blue的循环疲劳抗力高于原厂状态Reciproc(P < .05)。
- 用于根管预备的Reciproc Blue循环疲劳抗力高于用于根管预备的Reciproc(P < .05)。
- 有滑行通道与无滑行通道使用的Reciproc和Reciproc Blue之间循环疲劳抗力无统计学显著差异(P > .05)。
- 器械折断片段平均长度无统计学显著差异(P > .05)。
临床怎么用
- 在体外条件下,是否为Reciproc或Reciproc Blue锉创建滑行通道不影响其循环疲劳抗力。
- Reciproc Blue可能比Reciproc具有更高的循环疲劳抗力。
局限性
- 本研究为体外实验室研究,结果需结合临床情况谨慎解读。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The purpose of this article was to compare the cyclic fatigue resistance of Reciproc and Reciproc Blue files (VDW GmbH, Munich, Germany) that were used to prepare root canals of mandibular molar teeth with or without a glide path. METHODS: Sixty Reciproc R25 and 60 Reciproc Blue R25 files were used. The Reciproc and Reciproc Blue groups were divided into 3 subgroups (ie, as received condition, used without a glide path, and used with a glide path). All the instruments were rotated in a stainless steel artificial canal with an inner diameter of 1.5 mm, a 60° angle of curvature, and a radius of curvature of 5 mm until fracture occurred. The number of cycle to fracture was calculated, and the length of the fractured segments was measured. The Kruskal-Wallis test was performed to statistically analyze the data using SPSS 21.0 software (IBM Corp, Armonk, NY) at a 5% significance level. RESULTS: The cyclic fatigue resistance of as received condition Reciproc Blue files was found to be higher than as received condition Reciproc files (P < .05). Reciproc Blue files used for root canal preparation showed higher cyclic fatigue resistance than Reciproc files used for root canal preparation (P < .05). There was no statistically significant difference between Reciproc and Reciproc Blue files used with a glide path and without a glide path (P > .05). There was no statistically significant difference in the mean length of the fractured fragments of the instruments (P > .05). CONCLUSIONS: Within the limitations of this in vitro study, it was concluded that creating a glide path using ProGlider files had no effect on the cyclic fatigue resistance of RPC and RPC Blue files.
引用
- Özyürek T, Uslu G, Yılmaz K, et al. Effect of Glide Path Creating on Cyclic Fatigue Resistance of Reciproc and Reciproc Blue Nickel-titanium Files: A Laboratory Study. J Endod. 2018. PMID: 29680720. doi:10.1016/j.joen.2018.03.004
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。