本研究通过体外实验探讨冲洗时间对EndoVac根尖负压冲洗系统抗菌效果的影响,并基于最佳时间进行体内验证。体外使用44颗离体前磨牙接种粪肠球菌21天,分别冲洗10秒及MicroCannula 15至90秒;体内对20例根尖周炎患者用8.25%次氯酸钠冲洗60秒。结果显示冲洗时间延长可提高抗菌效果,体内冲洗60秒后细菌培养均为阴性。

要点

  • EndoVac系统抗菌效果随冲洗时间增加而提高。
  • 体内使用8.25%次氯酸钠经MicroCannula冲洗60秒可达到培养阴性。
  • 推荐每根管MicroCannula冲洗至少60秒以优化细菌清除。

核心结论

  • 经EndoVac根尖负压冲洗装置输送8.25%次氯酸钠60秒可能实现微生物消除。

研究设计与样本

  • 体外实验使用44颗离体前磨牙接种粪肠球菌21天,其中4颗为阳性对照,40颗分别冲洗不同时间(每组8颗)。
  • 体内验证对20例根尖周炎患者使用8.25%次氯酸钠经EndoVac MicroCannula冲洗60秒。

主要发现

  • 与对照组相比,冲洗显著降低细菌数量(p < .05)。
  • EndoVac冲洗的抗菌效果随冲洗时间增加而提高(p < .05)。
  • 体内冲洗60秒后采集的样本全部培养阴性。

关键数值与推荐

  • 体外使用2.5%次氯酸钠,先经MacroCannula冲洗10秒,再经MicroCannula分别冲洗15、30、45、60或90秒。
  • 体内使用8.25%次氯酸钠经MicroCannula冲洗60秒。

临床怎么用

  • 使用EndoVac系统时,次氯酸钠经MicroCannula每根管冲洗至少60秒以达到最佳根管内细菌清除。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The effect of irrigation time on the antimicrobial efficacy of an apical negative pressure irrigation system was examined in vitro, followed by validation of the antimicrobial effect in vivo using the identified optimal irrigation time. METHODS: For the in vitro experiment, 44 extracted premolars were decoronated, instrumented, autoclaved and inoculated with Enterococcus faecalis (ATCC 29212) for 21 days. Four teeth were used as positive control, without irrigation. Each of the remaining 40 teeth was irrigated with 2.5% NaOCl, delivered via the EndoVac MacroCannula for 10 s, and subsequently via the EndoVac MicroCannula for 15, 30, 45, 60 or 90 s per canal, respectively (N = 8). After irrigation, microbial samples were collected, transferred to BHI broth and incubated for counting of bacterial colony forming units (CFUs). Based on the in vitro results, 8.25% NaOCl was delivered via the EndoVac MicroCannula for 60 s, during root canal treatment of 20 human subjects presented with apical periodontitis. Microbial samples retrieved in vivo prior to canal instrumentation (S0), after chemomechanical debridement (S1) and after irrigation with EndoVac (S2) were cultured in an anaerobic chamber for 7 days for CFU evaluation. RESULTS: Compared with the control, irrigation significantly reduced bacterial populations (p < .05). Irrigation delivery via the EndoVac demonstrated improved antibacterial efficacy with increased irrigation time (p < .05). Samples retrieved from canals after NaOCl delivery in vivo with the EndoVac for 60 s were all culture-negative. CONCLUSIONS: Microbial elimination may be achieved with 8.25% NaOCl delivered via the EndoVac apical negative pressure irrigation device for 60 s. CLINICAL SIGNIFICANCE: With the use of the EndoVac apical negative pressure irrigant delivery system, optimal elimination of the intracanal bacterial load can only be achieved when sodium hypochlorite is delivered via the MicroCannula for at least 60 s per canal.