管电流设置对CBCT诊断牙根折裂的影响:体外研究
结果显示管电流对诊断有显著影响,8.0和10.0 mA的AUROC显著更高(P ≤ .0389),两者间无统计学差异。
要点
- 管电流设置显著影响CBCT对牙根折裂的诊断检测能力。
- 8.0和10.0 mA的诊断判别能力显著优于4.0和5.0 mA。
- 4.0和5.0 mA虽可接受诊断,但判别能力较低。
核心结论
- 管电流对CBCT图像中牙根折裂的诊断检测有显著影响。
- 4.0和5.0 mA虽可接受诊断牙根折裂,但判别能力低于8.0和10.0 mA。
研究设计与样本
- 牙齿分为对照组和折裂组,折裂组使用万能机械测试机诱导根折后置于体模中。
- 使用Scanora 3DX设备在4.0、5.0、6.3、8.0和10.0 mA五种管电流下采集CBCT图像。
- 两名检查者(E1和E2)按5点置信度量表分类图像。
主要发现
- 检查者内可重复性为显著一致(κE1 = 0.791,κE2 = 0.695),检查者间为中等一致(κE1×E2 = 0.545)。
- 8.0和10.0 mA的AUROC显著更高(P ≤ .0389),且这两种管电流设置间无统计学差异。
关键数值与推荐
- 诊断性能评估采用受试者工作特征曲线下面积(AUROC)。
- 可重复性评估采用kappa统计量。
- 8.0和10.0 mA的AUROC显著高于其他设置(P ≤ .0389)。
临床怎么用
- 在CBCT检测牙根折裂时,可优先考虑使用8.0或10.0 mA管电流设置以提高诊断判别能力。
- 使用4.0或5.0 mA时虽可诊断,但需注意其判别能力较低。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study assessed the influence of tube current settings (milliamperes [mA]) on the diagnostic detection of root fractures (RFs) using cone-beam computed tomographic (CBCT) imaging. METHOD: Sixty-eight human anterior and posterior teeth were submitted to root canal preparation, and 34 root canals were filled. The teeth were divided into 2 groups: the control group and the fractured group. RFs were induced using a universal mechanical testing machine; afterward, the teeth were placed in a phantom. Images were acquired using a Scanora 3DX unit (Soredex, Tuusula, Finland) with 5 different mA settings: 4.0, 5.0, 6.3, 8.0, and 10.0. Two examiners (E1 and E2) classified the images according to a 5-point confidence scale. Intra- and interexaminer reproducibility was assessed using the kappa statistic; diagnostic performance was assessed using the area under the receiver operating characteristic curve (AUROC). RESULTS: Intra- and interexaminer reproducibility showed substantial (κE1 = 0.791 and κE2 = 0.695) and moderate (κE1 × E2 = 0.545) agreement, respectively. AUROC was significantly higher (P ≤ .0389) at 8.0 and 10.0 mA and showed no statistical difference between the 2 tube current settings. CONCLUSIONS: Tube current has a significant influence on the diagnostic detection of RFs in CBCT images. Despite the acceptable diagnosis of RFs using 4.0 and 5.0 mA, those settings had lower discrimination abilities when compared with settings of 8.0 and 10.0 mA.
引用
- Tangari-Meira R, Vancetto JR, Dovigo LN, et al. Influence of Tube Current Settings on Diagnostic Detection of Root Fractures Using Cone-beam Computed Tomography: An In Vitro Study. J Endod. 2017. PMID: 28818444. doi:10.1016/j.joen.2017.05.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。