5种根尖定位仪在不同操作方案下的准确性评价
结果显示在根尖孔处测量时平均误差最小,电子回退未带来额外优势。
要点
- 在根尖孔处测量时5种EAL平均误差均较小
- 电子回退方案未优于直接到达根尖孔
- 不同EAL间仅在-1.0与over/0.0方案存在差异
核心结论
- 在根尖孔处进行电子根管长度测量时结果最佳,电子回退较到达根尖孔无额外优势。
主要发现
- 0.0方案平均误差为RZX 0.10 mm、RAY 0.13 mm、AID 0.16 mm、PRO 0.23 mm、PIXI 0.10 mm。
- 0.0与over/0.0方案间差异无统计学意义(P > .05)。
- 0.0与-1.0、0.0/-1.0比较,多数EAL差异有统计学意义(P < .05)。
- EAL间比较仅在-1.0和over/0.0方案差异有统计学意义(P < .05)。
关键数值与推荐
- 方案-1.0:插入至根尖孔下方1.0 mm。
- 方案0.0/-1.0:插入至根尖孔后回退1.0 mm。
- 方案0.0:插入至根尖孔。
- 方案over/0.0:插入至「over」后回退至根尖孔。
临床怎么用
- 临床使用EAL时宜以到达根尖孔的读数作为工作长度参考,不必依赖电子回退。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The aim of this study was to evaluate the accuracy of 5 electronic apex locators (EALs): Root ZX II (RZX; J Morita, Tokyo, Japan), Raypex 6 (RAY; VDW GmbH, Munich, Germany), Apex ID (AID; SybronEndo, Orange, CA), Propex II (PRO; Dentsply Maillefer, Ballaigues, Switzerland), and Propex Pixi (PIXI, Dentsply Maillefer) when used in the following protocols: (1) -1.0, insertion up to 1.0 mm below the apical foramen (AF); (2) 0.0/-1.0, insertion until the AF and withdrawn 1.0 mm short of the AF; (3) 0.0, insertion until the AF; and (4) over/0.0, insertion until "over" and withdrawal to AF. METHODS: Thirty human lower premolars had coronary accesses and cervical and middle thirds preparations performed, allowing AF standardization (200 μm). Using an alginate experimental model, root canal length (RCL) measurements were performed sequentially with EALs following each of the protocols. RESULTS: Considering the suggested protocols, the lowest mean error values were observed in 0.0, 0.10 mm (RZX), 0.13 mm (RAY), 0.16 mm (AID), 0.23 mm (PRO), and 0.10 mm (PIXI), without a significant difference for over/0.0 (P > .05). Comparing the results obtained in 0.0 with those found in -1.0 and 0.0/-1.0, significant differences were observed for most EALs (P < .05). For the comparison between EALs, significant differences were observed only in protocols -1.0 and over/0.0 (P < .05). CONCLUSIONS: Under the conditions of the study, it was concluded that, regardless of the mechanism of the device, the best results were found when electronic RCL measurement was performed at the AF; furthermore, the electronic withdrawal did not offer any additional advantage over the reach of the AF.
引用
- Oliveira TN, Vivacqua-Gomes N, Bernardes RA, et al. Determination of the Accuracy of 5 Electronic Apex Locators in the Function of Different Employment Protocols. J Endod. 2017. PMID: 28693895. doi:10.1016/j.joen.2017.03.039
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。