不同锥形束CT设备对充填根管体积伪影的评估
结果显示CBCT与micro-CT体积差异显著(P < .05),且不同CBCT设备间也有显著差异,Promax 3D Max测得的根管体积显著大于其他CBCT设备(P < .05)。
要点
- CBCT设备比micro-CT显示更多根管体积失真伪影。
- 相同体素大小下不同CBCT设备测得的根管体积存在显著差异。
- Promax 3D Max测得的根管体积显著大于NewTom VGi evo和3D Accuitomo 170。
核心结论
- CBCT设备比micro-CT成像显示更多的根管体积失真伪影。
- 在相同体素大小扫描时,不同CBCT设备测得的根管体积存在差异。
研究设计与样本
- 所有牙齿用牙胶和AH Plus封闭剂以单尖法充填,分别用三种CBCT设备(Promax 3D Max、NewTom VGi evo、3D Accuitomo 170)以相同体素大小(0.2 mm3)扫描,并与micro-CT参考标准比较。
主要发现
- micro-CT与CBCT图像之间在体积失真方面存在显著差异(P < .05)。
- 不同CBCT设备之间也存在显著差异。
- Promax 3D Max测得的根管体积显著大于其他CBCT设备(P < .05)。
- NewTom VGi evo与3D Accuitomo 170结果相似,无显著差异(P > .05)。
临床怎么用
- 临床评估充填根管体积时需注意CBCT设备可能引入的体积失真伪影。
- 不同CBCT设备在相同体素大小下测得的根管体积可能不同,选择设备时应考虑此差异。
局限性
- 需要进一步研究不同封闭剂材料对CBCT成像的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Artifacts in cone-beam computed tomographic (CBCT) imaging may compromise radiodiagnosis. Obturation materials for endodontic treatment may present with variable material density and thus also cause distinct artifact expression. The aim of this study was to assess the volume distortion artifact of root canal sealers using CBCT devices and micro-CT imaging as a reference. METHODS: Thirty single-root mandibular central incisors were used for this study. Teeth were prepared with EndoSequence rotary nickel-titanium files (Brasseler USA, Savannah, GA) and divided into 3 groups. Canals were obturated with gutta-percha and AH Plus root canal sealer (Dentsply Maillefer, Ballaigues, Switzerland) using single-cone filling techniques. Each tooth was scanned with different CBCT devices (ie, Promax 3D Max [Planmeca Inc, Roselle, IL], NewTom VGi evo [NewTom, Verona, Italy], and 3D Accuitomo 170 [J Morita, Kyoto, Japan]) with the same voxel size (0.2 mm3) and compared with micro-CT imaging as a reference standard. RESULTS: The results showed a significant difference in terms of volume distortion between micro-CT and CBCT images (P < .05). There were also significant differences among CBCT devices. Promax 3D Max measurements showed significantly larger root canal volumes than the other CBCT machines (P < .05). However, NewTom VGi evo and 3D Accuitomo 170 showed similar results without any significant difference (P > .05). CONCLUSIONS: CBCT devices showed more volumetric distortion artifact than micro-CT imaging. The volume was variable for different CBCT devices while scanning at the same voxel size. However, to assess the effect of sealer materials on CBCT imaging, further studies should be conducted for different sealers.
引用
- Celikten B, Jacobs R, deFaria Vasconcelos K, et al. Assessment of Volumetric Distortion Artifact in Filled Root Canals Using Different Cone-beam Computed Tomographic Devices. J Endod. 2017. PMID: 28673491. doi:10.1016/j.joen.2017.03.035
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。