四种单支镍钛器械的循环疲劳抗力比较
本研究比较Reciproc Blue、HyFlex EDM、WaveOne Gold和OneShape四种单支镍钛系统的循环疲劳抗力。每种器械30支在模拟根管中旋转至断裂并记录时间。结果显示HyFlex EDM疲劳抗力最高,OneShape最低;Reciproc Blue高于WaveOne Gold;断裂片段长度无显著差异。
要点
- HyFlex EDM循环疲劳抗力最高,OneShape最低
- Reciproc Blue疲劳抗力显著高于WaveOne Gold
- 四种器械断裂片段长度无统计学差异
核心结论
- 在本体外研究限制内,HyFlex EDM的循环疲劳抗力高于OneShape、Reciproc Blue和WaveOne Gold。
研究设计与样本
- 体外研究,每组30支单支镍钛器械,在模拟根管中旋转至断裂并记录时间。
主要发现
- HyFlex EDM断裂时间(3456.33 ± 633.37秒)最高,OneShape(1221.63 ± 812.4秒)最低(P < .05)。
- Reciproc Blue平均断裂循环数(2875.89 ± 105.35)显著高于WaveOne Gold(1737.00 ± 376.32)(P < .05)。
- 断裂片段平均长度无统计学显著差异(P > .05)。
关键数值与推荐
- 模拟根管内径1.5 mm,弯曲角度60°,弯曲半径5 mm。
- 统计分析采用Kruskal-Wallis和post hoc Dunn检验,显著性水平5%。
临床怎么用
- 选择单支镍钛器械时可参考其循环疲劳抗力,HyFlex EDM抗力较高。
- 在弯曲根管中,OneShape的循环疲劳抗力相对较低,需注意使用次数。
局限性
- 本研究为体外研究,结果不能直接外推至临床。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The purpose of this study was to compare the cyclic fatigue resistances of Reciproc Blue (VDW, Munich, Germany), HyFlex EDM (Coltene/Whaledent, Altstätten, Switzerland), WaveOne Gold (Dentsply Maillefer, Ballaigues, Switzerland), and OneShape (Micro Mega, Besancon, France) single-file NiTi systems. METHODS: Thirty Reciproc Blue R25 (25/.08), 30 HyFlex EDM (25/.∼), 30 WaveOne Gold Primary (25/.07), and 30 OneShape (25/.06) instruments were included in this study. All the instruments were rotated in artificial canals, which were made of stainless steel with an inner diameter of 1.5 mm, a 60° angle of curvature, and radii of curvatures of 5 mm until fracture occurred, and the time to fracture was recorded in seconds using a digital chronometer. The data were analyzed statistically using Kruskal-Wallis and post hoc Dunn tests via SPSS 21.0 software (SPSS Inc, Chicago, IL). The statistical significance level was set at 5%. RESULTS: The HyFlex EDM file (3456.33 ± 633.37) file had the statistically highest fatigue resistance, and the OneShape file (1221.63 ± 812.4) had the least fatigue resistance (P < .05). The mean number of cycles to fracture of the Reciproc Blue file (2875.89 ± 105.35) file was statistically higher than the WaveOne Gold file (1737.00 ± 376.32) (P < .05). There was no statistically significant difference (P > .05) in the mean length of the fractured fragments of the files (P > .05). CONCLUSION: Within the limitations of the present in vitro study, it was found that cyclic fatigue resistance of HyFlex EDM files was higher than the cyclic fatigue resistances of OneShape, Reciproc Blue, and WaveOne Gold files.
引用
- Gündoğar M, Özyürek T. Cyclic Fatigue Resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue Nickel-titanium Instruments. J Endod. 2017. PMID: 28527845. doi:10.1016/j.joen.2017.03.009