使用谱域和扫频源光学相干断层扫描检测根尖牙根裂纹
本研究选取20颗拔除的人下颌单根前牙,根管预备后以微计算机断层扫描(micro-CT)为金标准识别根尖微裂纹,再用谱域OCT(SD-OCT)和扫频源OCT成像并由3名独立检查者盲法评估。micro-CT显示12颗(60%)牙根存在根尖牙本质微裂纹,两种OCT均能在对应位置显示微裂纹线,SD-OCT诊断性能较优但两种设备间无统计学显著差异(P > .05)。作者认为两种OCT系统均有望用于根尖微裂纹的诊断。
要点
- micro-CT显示12颗(60%)牙根存在根尖牙本质微裂纹。
- 两种OCT系统均能在与micro-CT对应截面相同位置显示微裂纹线。
- SD-OCT诊断性能较优,但与扫频源OCT间无统计学显著差异(P > .05)。
核心结论
- 两种OCT系统均被验证具有检测根尖微裂纹的能力,作者认为它们有望成为根尖微裂纹诊断的工具。
研究设计与样本
- 选取20颗拔除的人下颌单根前牙,使用R40 Reciproc锉(VDW,慕尼黑,德国)行根管预备。
- 以micro-CT扫描识别根尖微裂纹作为金标准,随后对根尖部分行SD-OCT和扫频源OCT成像。
- 由3名独立检查者盲法评估图像以检测微裂纹,计算各OCT设备的诊断性能并进行统计分析。
主要发现
- 基于micro-CT图像,12颗(60%)牙根在根尖区域存在牙本质微裂纹。
- 两种OCT系统生成的图像能够在与相应micro-CT横截面相同的位置显示微裂纹线。
- SD-OCT设备的诊断性能较优,但两种OCT设备之间无统计学显著差异(P > .05)。
- 检查者间一致性:SD-OCT系统为较大至几乎完全一致,扫频源OCT系统为中等至几乎完全一致。
- 检查者内一致性:两种OCT设备均为较大至几乎完全一致。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study aimed to evaluate the ability of 2 optical coherence tomographic (OCT) systems to detect apical dentinal microcracks. METHODS: Twenty extracted human single-rooted mandibular incisors were selected. After root canal preparation with an R40 Reciproc file (VDW, Munich, Germany), the specimens presenting apical microcracks were identified using micro-computed tomographic (micro-CT) scanning as the gold standard. Then, the apical portions of the roots were imaged with spectral domain OCT (SD-OCT) and swept-source OCT systems, and the resulting images were blindly evaluated by 3 independent examiners to detect microcracks. The diagnostic performance of each OCT device was calculated, and statistical analysis was performed. RESULTS: Based on the micro-CT images, 12 (60%) roots presented dentinal microcracks in the apical region. The images generated by the OCT systems were able to show microcrack lines at the same location as the corresponding micro-CT cross sections. Although the diagnostic performance of the SD-OCT device was superior, there were no statistically significant differences between the 2 OCT devices (P > .05). Interexaminer agreement was substantial to almost perfect for the SD-OCT system and moderate to almost perfect for the swept-source OCT system, whereas intraexaminer agreement was substantial to almost perfect for both OCT devices. CONCLUSIONS: The detection ability verified for both OCT systems renders them promising tools for the diagnosis of apical microcracks.
引用
- de Oliveira BP, Câmara AC, Duarte DA, et al. Detection of Apical Root Cracks Using Spectral Domain and Swept-source Optical Coherence Tomography. J Endod. 2017. PMID: 28416309. doi:10.1016/j.joen.2017.01.019