影响口内扫描边缘质量的混杂因素
研究评估邻牙存在、牙龈邻近度及口内扫描杆操作受阻等临床因素对冠边缘口内扫描质量的影响。在牙颌模型上预备下颌磨牙全瓷冠,使用口内扫描仪在不同条件下获取40次扫描,以边缘曲率均值评估质量。结果显示颊侧平龈边缘、邻牙存在及牙龈邻近度显著影响部分边缘区域的曲率。
要点
- 舌侧边缘曲率不受扫描条件影响。
- 颊侧平龈边缘曲率受扫描条件显著影响。
- 远中边缘曲率受全部三种混杂因素显著影响。
核心结论
- 商用口内扫描仪记录的边缘曲率(锐度)受遮挡视野的临床因素显著影响。
研究设计与样本
- 在牙颌模型上预备下颌磨牙全瓷冠,使用口内扫描仪在8组不同条件下各获取5次扫描,共40次扫描,以边缘曲率均值为指标分析混杂因素影响。
主要发现
- 舌侧边缘曲率在所有扫描条件下保持一致。
- 颊侧边缘位于平龈时曲率受显著影响。
- 近中边缘曲率在邻牙存在和牙龈邻近时受显著影响。
- 远中边缘曲率受全部三种混杂因素显著影响。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To assess the effect of clinical factors on the quality of intra-oral scans of crown margins. These factors are; presence of adjacent teeth, proximity to gingivae, encumbrance of wand positioning within oral cavity. METHODS: A typodont lower molar (Frasaco, Germany) was prepared for an all-ceramic crown with 1.5mm supragingival (lingual) and equigingival (buccal) margins. The tooth was scanned in a model scanner, creating a master scan. An intra-oral scanner (IOS) (Omnicam, Sirona Dental) was used to acquire sets of 5 scans each, under varying conditions; (1) the presence/absence of adjacent teeth, (2) model mounted in manikin head/hand-held, (3) with/without a 1mm shim to elevate the margin. Every combination was investigated, yielding 40 scans (8 groups of 5). The master scan margin was identified by selecting the highest curvature region (>1.8). The master was aligned to each IOS scan, and 4 regions of each IOS scan margin were extracted, lying within 100μm of predefined mesial, distal, buccal and lingual sections of the master margin. The mean curvature of each margin section was calculated using Meshlab. The effect of each confounding factor on margin curvature was analysed using ANOVA. RESULTS: Lingual margin curvature remained consistent regardless of scanning conditions. Buccal margin curvature was significantly affected when located equigingivally. Mesial margin curvature was significantly affected in the presence of adjacent teeth and proximity to the gingivae. Distal margin curvature was significantly affected by all three confounding factors. CONCLUSIONS: The curvature (sharpness) of the margin recorded by a commercial IOS is significantly affected by clinical factors obscuring visibility.
引用
- Keeling A, Wu J, Ferrari M, et al. Confounding factors affecting the marginal quality of an intra-oral scan. J Dent. 2017. PMID: 28189718. doi:10.1016/j.jdent.2017.02.003
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。