不同冲洗液对两种电子根尖定位仪定位模拟性根穿孔准确性的影响
本研究用20颗离体单根人牙在根中段人工穿孔,评估Dentaport ZX与Rootor两种电子根尖定位仪在干燥条件及2.5%次氯酸钠、0.9%氯化钠、17%EDTA冲洗液存在下定位穿孔的准确性。两种定位仪在干燥条件下测量最准确;Dentaport ZX在不同冲洗液存在下比Rootor更准确,根管内容物影响两种定位仪的准确性。
要点
- 两种电子根尖定位仪在干燥根管中定位模拟性根穿孔最准确。
- Dentaport ZX在不同冲洗液存在下定位穿孔的准确性优于Rootor。
- 根管内容物会影响两种电子根尖定位仪的准确性。
核心结论
- Dentaport ZX在不同冲洗液存在下定位根穿孔的准确性优于Rootor。
- 根管内容物影响两种电子根尖定位仪的准确性,干燥根管中测量最准确。
研究设计与样本
- 离体研究:20颗拔除的单根人牙在根中段人工穿孔,包埋于藻酸盐模型中。
- 用20号K锉分别在干燥条件及2.5%次氯酸钠、0.9%氯化钠、17%EDTA存在下由两种电子根尖定位仪测量穿孔位置。
- 统计采用Friedman检验与Wilcoxon符号秩检验,显著性水平P < .05。
主要发现
- 两种电子根尖定位仪均在干燥条件下获得最准确的测量结果。
- Dentaport ZX组中EDTA测量最准确、次氯酸钠最不准确;Rootor组中氯化钠测量比EDTA更接近实际根管长度。
- 在次氯酸钠、氯化钠和EDTA测量条件下,两种电子根尖定位仪间差异有统计学意义(P < .05)。
关键数值与推荐
- 以实际根管长度至穿孔处为参照,从穿孔的电子测量长度中减去实际根管长度进行统计分析。
- 显著性水平为P < .05。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The aim of this study was to assess the accuracy of the Dentaport ZX (Morita Co, Kyoto, Japan) and the Rootor (Meta Biomed, Cheongwon-gun, Korea) electronic apex locators (EALs) in detecting root perforations in dry conditions and in the presence of the following irrigation solutions: 2.5% sodium hypochlorite (NaOCl), 0.9% saline solution (NaCl), and 17% EDTA. METHODS: Twenty extracted, single-rooted human teeth were perforated artificially in the middle section. The actual canal lengths (ALs) up to the perforation site were determined, and then the teeth were embedded in an alginate mold. The electronic measurements of the perforations were obtained using a size 20 K-file by each EAL in various conditions. For each tooth, the AL was subtracted from the electronic length of the perforation. Statistical analyses were performed using the Friedman and Wilcoxon signed rank tests at a significance level of P < .05. RESULTS: The most accurate measurements for both EALs were obtained in dry conditions. Among the irrigation solutions in the Dentaport ZX group, EDTA gave the most accurate results, and NaOCl gave the least accurate ones. However, measurements with NaCl were closer to the AL than those obtained with EDTA for the Rootor group. Significant differences were noted among the EALs when the measurements were taken with NaOCl, NaCl, and EDTA (P < .05). CONCLUSIONS: The Dentaport ZX was more accurate compared with the Rootor in the presence of different irrigants. The content of the root canal affected the accuracy of both EALs. The most accurate measurements were obtained in dry canals.
引用
- Altunbaş D, Kuştarcı A, Toyoğlu M, et al. The Influence of Various Irrigants on the Accuracy of 2 Electronic Apex Locators in Locating Simulated Root Perforations. J Endod. 2017. PMID: 28104321. doi:10.1016/j.joen.2016.10.031
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。