文献速递口腔颌面外科牙周病

多波长半导体激光用于口腔软组织切开与牙龈切除的多中心随机非劣效试验

[Evaluation of the efficacy and safety of multi-wavelength semiconductor laser for oral soft tissue incision and gingivectomy: a prospective, multicenter, randomized, parallel-controlled, non-inferiority clinical trial].
PubMed 文献 · Zhonghua Kou Qiang Yi Xue Za Zhi · 2026-09-01

该研究纳入156例需拔除下颌阻生智齿并行口腔软组织切开或牙龈切除术的患者,随机分为多波长半导体激光试验组和单波长半导体激光对照组,每组78例。结果显示,试验组术中止血和术后1周伤口愈合优良率均非劣于对照组,激光切割效率更高,两组不良事件发生率差异无统计学意义。结论认为多波长半导体激光治疗仪有效性和安全性非劣于对照产品。

要点

  • 多波长半导体激光术中止血优良率为100.0%,非劣于单波长激光。
  • 试验组激光切割效率(0.08±0.07)mm/s显著高于对照组(0.06±0.04)mm/s,P=0.004。
  • 两组不良事件发生率差异无统计学意义,试验组无不良事件发生。

核心结论

  • 多波长半导体激光治疗仪用于口腔软组织切开及牙龈切除术的有效性和安全性非劣于单波长半导体激光治疗仪,符合临床验证标准。

研究设计与样本

  • 前瞻性、多中心、随机、平行对照、非劣效临床试验,纳入2023年6月至2024年3月三家医院需拔除下颌阻生智齿并行口腔软组织切开或牙龈切除的患者156例,分层区组随机分为试验组78例和对照组78例。
  • 试验组使用多波长半导体激光治疗仪完成术区软组织切开及牙龈切除,对照组使用单波长半导体激光治疗仪完成相应操作,术后1周和1个月随访。

主要发现

  • 试验组术中止血优良率为100.0%(78/78),术后1周伤口愈合优良率为100.0%(78/78),均非劣于对照组的100.0%(78/78)和98.7%(77/78)。
  • 试验组激光切割效率为(0.08±0.07)mm/s,显著高于对照组的(0.06±0.04)mm/s(Z=-2.87,P=0.004)。
  • 两组在消毒便利性、操作便利性、操作安全性及稳定性方面差异均无统计学意义(均P>0.05)。
  • 安全性方面,试验组无不良事件发生,对照组发生轻、中度不良事件2例(2.6%)和严重不良事件1例(1.3%),两组不良事件发生率差异无统计学意义(P>0.05)。

临床怎么用

  • 多波长半导体激光可作为口腔软组织切开和牙龈切除的备选工具,其止血和愈合效果与单波长激光相当,且切割效率更高。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Objective: To evaluate the clinical efficacy and safety of the multi-wavelength semiconductor laser therapeutic instrument in oral soft tissue incision and gingivectomy. Methods: A total of 156 patients who received mandibular impacted wisdom tooth extraction combined with oral soft tissue incision or gingivectomy at the Stomatological Hospital of Zhejiang University School of Medicine, Zhejiang Hospital and Wenzhou Central Hospital between June 2023 and March 2024 were enrolled and divided into experimental group (n=78) and control group (n=78) by stratified block randomization method. During the operation, the experimental group used the multi-wavelength semiconductor laser therapeutic instrument to complete soft tissue incision and gingivectomy in the surgical area, while the control group used the single-wavelength semiconductor laser therapeutic instrument for the corresponding procedures. Follow-up assessments were conducted one week and one month postoperatively for both groups of patients. Primary outcomes were the excellent and good rate of intraoperative hemostasis and wound healing at one week postoperatively, which were analyzed by non-inferiority test. Secondary outcomes included laser cutting efficiency, convenience of disinfection, operational convenience, operational safety and stability, which were assessed using the Wilcoxon rank-sum test, χ2 test or Fisher's exact probability method. Safety indicator was the incidence of adverse events, which was analyzed using Fisher's exact probability method. Results: Primary efficacy evaluation: the excellent and good rate of intraoperative hemostasis [100.0% (78/78) vs. 100.0% (78/78)] and wound healing [100.0% (78/78) vs 98.7% (77/78)] in the experimental group were both non-inferior to those in the control group. Secondary efficacy evaluation: the laser cutting efficiency in experimental group [(0.08±0.07) mm/s] was significantly higher than that in control group [(0.06±0.04) mm/s] (Z=-2.87, P=0.004). There were no significant differences in disinfection convenience, operational convenience, safety, and stability between the two groups (all P>0.05). Regarding safety evaluation, no adverse events occurred in the experimental group, while 2 cases (2.6%) of mild or moderate adverse events and 1 case (1.3%) of serious adverse events occurred in the control group, with no significant difference (P>0.05). Conclusions: The efficacy and safety of the multi-wavelength semiconductor laser therapeutic instrument are non-inferior to the control products, meeting the standards for clinical validation. 目的: 评价多波长半导体激光治疗仪用于口腔软组织切开、牙龈切除术的临床有效性和安全性。 方法: 纳入2023年6月至2024年3月在浙江大学医学院附属口腔医院、浙江医院、温州市中心医院就诊,需拔除下颌阻生智齿且需进行口腔软组织切开、牙龈切除的156例患者,采用分层区组随机法将患者分为试验组(78例)和对照组(78例),术中试验组使用多波长半导体激光治疗仪完成术区软组织切开及牙龈切除操作,对照组使用单波长半导体激光治疗仪完成术区软组织切开及牙龈切除操作。对两组患者术后1周、1个月进行随访。主要有效性评价指标为术中止血效果优良率、术后1周伤口愈合情况优良率,采用非劣效检验;次要有效性评价指标为激光切割效率、设备的消毒便利性、操作便利性、操作安全性及稳定性,采用Wilcoxon秩和检验、χ2检验或Fisher精确概率法检验;安全性指标为不良事件发生率,采用Fisher精确概率法检验。 结果: 主要有效性评价:试验组术中激光止血效果优良率为100.0%(78/78),术后1周伤口愈合情况优良率为100.0%(78/78),均非劣于对照组的100.0%(78/78)、98.7%(77/78)。次要有效性评价:试验组激光切割效率[(0.08±0.07)mm/s]显著高于对照组[(0.06±0.04)mm/s](Z=-2.87,P=0.004),消毒便利性、操作便利性、安全性及稳定性评价与对照组相比差异均无统计学意义(均P>0.05)。安全性评价方面,试验组无不良事件发生,对照组发生3例不良事件,2例(2.6%)为轻、中度不良事件,1例(1.3%)为严重不良事件,两组不良事件发生率差异无统计学意义(P>0.05)。 结论: 多波长半导体激光治疗仪用于下颌阻生智齿拔除患者口腔软组织切开、牙龈切除术的有效性和安全性良好,与对照组相当。.

来源:Zhonghua Kou Qiang Yi Xue Za Zhi,查看原文。版权归原作者所有。

引用

  1. Chen XY, Hao YL, Han CY, et al. [Evaluation of the efficacy and safety of multi-wavelength semiconductor laser for oral soft tissue incision and gingivectomy: a prospective, multicenter, randomized, parallel-controlled, non-inferiority clinical trial]. Zhonghua Kou Qiang Yi Xue Za Zhi. 2026. PMID: 42677436. doi:10.3760/cma.j.cn112144-20260303-00147

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。