口内超声牙周测量准确性与可靠性

Periodontal Measurement Accuracy and Reliability Using Intraoral Ultrasound.
PubMed 文献 · J Dent Res · 2026-07-18

本研究评估口内超声(iUS)测量牙周参数的准确性与可靠性:离体用58颗牙以显微CT为金标准,临床在19名正畸青少年中由3名评估者测量134张图像;iUS对牙槽骨水平与厚度准确性高,体内重复测量信度良好至优秀。

要点

  • iUS测量牙槽骨水平与厚度与显微CT高度一致,无显著差异。
  • 标准化解剖引导协议使iUS体内测量重复性与评估者间信度良好至优秀。
  • 结论需在中重度牙周病患者中进一步验证后方可用于纵向监测。

核心结论

  • iUS在尸体与健康青少年中提供高度可重复且可靠的牙周测量。

研究设计与样本

  • 离体准确性评估:3具尸体标本共58颗牙,以显微CT为金标准,2名评估者在50张iUS图像上重复测量牙槽骨水平(ABL)与牙槽骨厚度(ABT)。
  • 体内临床可靠性评估:19名接受正畸治疗的青少年,3名评估者在134张iUS图像上测量ABL、ABT、附着龈厚度与游离龈厚度。
  • 采用20 MHz定制口内超声系统(旋转探头)及标准化解剖引导协议进行成像与测量。

主要发现

  • 离体ABL测量平均绝对差(MAD)0.18-0.31 mm,组内相关系数(ICC)0.94-0.96。
  • 离体ABT测量MAD 0.05 mm,ICC 0.87。
  • iUS与显微CT测量ABL和ABT的配对t检验均无显著差异(所有P > 0.05)。
  • Bland-Altman分析显示ABL与ABT偏倚接近零,各评估者95%一致性界限窄。
  • 体内所有参数评估者内信度ICC ≥ 0.84、MAD ≤ 0.10 mm;评估者间信度ICC ≥ 0.82、MAD ≤ 0.18 mm。

关键数值与推荐

  • 离体ABL:MAD 0.18-0.31 mm,ICC 0.94-0.96。
  • 离体ABT:MAD 0.05 mm,ICC 0.87。
  • 体内评估者内:ICC ≥ 0.84,MAD ≤ 0.10 mm。
  • 体内评估者间:ICC ≥ 0.82,MAD ≤ 0.18 mm。

临床怎么用

  • 采用标准化解剖引导协议进行iUS成像与测量可降低操作者依赖性并提高可重复性。
  • iUS可作为无辐射、实时可视化软硬组织的牙周评估工具,但当前证据限于尸体与健康青少年。

局限性

  • 体内可靠性仅在19名接受正畸治疗的健康青少年中评估。
  • 需在中重度牙周病患者中进一步研究以充分验证iUS用于纵向监测。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Intraoral ultrasonography (iUS) has emerged as a noninvasive, radiation-free imaging modality capable of real-time visualization of both soft and hard tissues, yet its clinical accuracy and reliability remain incompletely characterized. This study aims to evaluate the accuracy and reliability of periodontal parameters measured on intraoral ultrasonograms. An in-house customized 20-MHz ultrasound imaging system with a rotational head was used to image the periodontium. Ex vivo accuracy was assessed in 58 teeth from 3 cadaver specimens with high-resolution micro-computed tomography (µCT) as the gold standard. Two raters independently performed repeated measurements of alveolar bone level (ABL) and alveolar bone thickness (ABT) on 50 iUS images using a standardized anatomy-guided protocol. Intermethod agreement was quantified using intraclass correlation coefficients (ICCs), mean absolute differences (MADs), paired t tests, and Bland-Altman analysis. In vivo clinical reliability was evaluated in 19 adolescents undergoing orthodontic treatment, with 3 raters measuring the ABL, ABT, attached gingival thickness, and free gingival thickness on 134 iUS images. Ex vivo analysis demonstrated excellent iUS accuracy and consistency for ABL (MAD 0.18-0.31 mm, ICCs 0.94-0.96) and excellent accuracy and good agreement for ABT (MAD 0.05 mm; ICCs 0.87). Paired t tests revealed no significant differences between iUS and µCT measurements for either parameter (all P > 0.05). Bland-Altman analysis demonstrated excellent agreement with near-zero bias for both ABL and ABT, with narrow 95% limits of agreement across raters. In vivo measurements exhibited outstanding intrarater reliability (ICCs ≥ 0.84; MAD ≤ 0.10 mm) and good-to-excellent interrater reliability (ICCs ≥ 0.82; MAD ≤ 0.18 mm) for all parameters. The standardized, anatomy-guided protocol minimized operator dependency, enabling reproducible assessments across raters. These results demonstrate that iUS provides highly reproducible and reliable periodontal measurements in cadavers and healthy adolescents. Further studies in patients with moderate-to-severe periodontal disease are required to fully validate iUS for longitudinal monitoring.

来源:J Dent Res,查看原文。版权归原作者所有。