微CT研究:三种镍钛系统在重度弯曲根管中的成形行为
结果显示三组在根管偏移和中心定位上无显著差异(P > .05),仅PTU在冠三分之一处的中心定位能力显著优于WOG(P = .035)。结论为三种系统整体成形效果相似,设计差异合理组合均可获得有效预备。
要点
- PTU、TRN和WOG在重度弯曲根管中的根管偏移无显著差异。
- PTU在冠三分之一的中心定位能力显著优于WOG(P = .035)。
- 不同设计和制造特征合理组合均可实现有效成形。
核心结论
- 三种镍钛系统在重度弯曲根管中的根管偏移相似,仅PTU在冠三分之一中心定位能力优于WOG。
研究设计与样本
- 预备前后进行微CT扫描,在距根尖3、5、7 mm处计算根管偏移和中心定位能力。
主要发现
- 三组在中心定位能力和根管偏移上无显著差异(P > .05)。
- PTU在冠三分之一处中心定位能力(均值0.595 mm)显著优于WOG(均值0.329 mm;P = .035)。
- 不同根管水平上的根管偏移组间无显著差异(P > .05)。
关键数值与推荐
- 根管弯曲度30°-40°;冲洗使用3%次氯酸钠,最终冲洗包括3%次氯酸钠、15% EDTA和次氯酸钠。
- PTU至F2(25/08),TRN Primary(26/04),WOG Primary(25/07)。
- 统计学采用方差分析和Bonferroni检验,P值设为0.05。
临床怎么用
- 对于重度弯曲根管,三种系统均可选用,PTU在冠三分之一中心定位更佳。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: To analyze the canal transportation and centering ability of rotary, ProTaper Ultimate (PTU) and TruNatomy (TRN), and reciprocating WaveOne Gold (WOG) instrumentation systems in curved root canals. METHODS: Forty-two upper human molars with curved mesiobuccal canals (30°-40°) were selected. The samples were assigned into 3 groups based on the instruments used: Group 1. PTU up to F2, 25/08; Group 2. TRN Primary file 26/04; Group 3. WOG Primary file 25/07. For irrigation, 3% sodium hypochlorite (NaOCl) was used throughout the instrumentation. Final irrigation protocol included 3% NaOCl, followed by 15% ethylenediaminetetraacetic acid and NaOCl. Each sample was scanned by micro-computed tomography (μCT) before and after the chemomechanical instrumentation. On μCT scans, the canal transportation and centring ability of the instrumentation systems were calculated at the levels of 3, 5, and 7 mm from the apex. Analysis of variance with Bonferroni test were used for the statistical analysis with the P value set at 0.05. RESULTS: There were no significant differences between the instrumentation systems regarding their centering ability and transportation (P > .05). PTU showed significantly better centering in the coronal third (mean 0.595 mm) compared with WOG (mean 0.329 mm; P = .035). Regarding canal transportation at different canal levels, no significant difference between the groups was found (P > .05). CONCLUSIONS: All tested instrumentation systems showed similar canal transportation, with one exception; PTU showed better centring ability in the coronal third. The study confirmed that different manufacturing and design features, if properly combined, may all result in valid operative sequence and proper shaping.
引用
- Bago I, Katić M, Milanović-Litre M, et al. Micro-CT Study of Shaping Behavior of ProTaper Ultimate, TruNatomy and Reciprocating WaveOne Gold in Severely Curved Root Canals. J Endod. 2026. PMID: 41887503. doi:10.1016/j.joen.2026.03.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。